2.2.10 MEASure
MEASure[:SCALar]:VOLTage[:DC]?
Queries the input voltage of the load.
Command Syntax: MEASure[:SCALar]:VOLTage[:DC]?
Parameter: None
Query Syntax: MEAS:VOLT?
Returned Parameter: <NR3>
Related Commands: MEAS:CURR?, MEAS:RES?, MEAS:POW?
MEASure[:SCALar]:CURRent[:DC]?
Queries the input current of the load.
Command Syntax: MEASure[:SCALar]:CURRent[:DC]?
Parameter: None
Query Syntax: MEAS:CURR?
Returned Parameter: <NR3>
Related Commands: MEAS:VOLT?, MEAS:RES?, MEAS:POW?
MEASure[:SCALar]:POWer[:DC]?
Queries the input power of the load.
Command Syntax: MEASure[:SCALar]:POWer[:DC]?
Parameter: None
Query Syntax: MEAS:POW?
Returned Parameter: <NR3>
Related Commands: MEAS:CURR?, MEAS:VOLT?, MEAS:RES?
MEASure[:SCALar]:RESistance[:DC]?
Queries the input resistance of the load.
Command Syntax: MEASure[:SCALar]:RESistance[:DC]?
Parameter: None
Query Syntax: MEAS:RES?
Returned Parameter: <NR3>
Related Commands: MEAS:CURR?, MEAS:VOLT?, MEAS:RES?
2.2.11 STATus Subsystem
STATus:QUEStionable[:EVENt]?
This query reads the Questionable Status Event register, and
returns a decimal value, which equals to the total weighted value
of all binary bits. Reading this register resets it to zero. See Status
Reporting for more details concerning the Questionable Status
Event register.
Query Syntax: STATus:QUEStionable[:EVENt]?
Parameter: None
Returned Parameter: <NR1>
STATus:QUEStionable:CONDition?
This query reads the Questionable Status Condition register. Each
0 to 1 transition of the condition of the Questionable Condition
register will set the corresponding bit of the Questionable Status
Event register. See Status Reporting for more details concerning
the Questionable Status Condition register.
Query Syntax: STATus:QUEStionable:CONDition?
Parameter: None
Returned Parameter: <NR1>
STATus:QUEStionable:ENABle
Sets the bits of the Questionable Status Enable register. The pro-
gramming parameters determine whether or not a 0 to 1 transition
in the Questionable Event register will set the QUES bit of the Sta-
tus Byte register. See Status Reporting for more details concern-
ing the Questionable Status Enable register.
Command Syntax: STATus:QUEStionable:ENABle <NRf>
Parameter: 0?65535
Power-on Value: refers to *PSC command
Example: STATus:QUEStionable:ENABle 64
Query Syntax: STATus:QUEStionable:ENABle?
Returned Parameter: <NR1>
Related Commands: *PSC
STATus:OPERation[:EVENt]?
This query reads the Operation Event register. Reading this regis-
ter resets it to zero. See Status Reporting for more details con-
cerning the Operation Event register.
Query Syntax: STATus:OPERation [:EVENt]?
Parameter: None
Returned Parameter: <NR1>
STATus:OPERation:CONDition?
This query reads the Operation Status Condition register. Each 0
to 1 transition of the condition of the Operation Condition register
will set the corresponding bit of the Operation Event register. See
Status Reporting for more details concerning the Operation Sta-
tus Condition register.
Query Syntax: STATus:OPERation:CONDition?
Parameter: None
Returned Parameter: <NR1>
STATus:OPERation:ENABle
Sets the bits of the Operation Status Enable register. The pro-
gramming parameters determine whether or not a 0 to 1 transition
in the Operation Event register will set the OPER bit of the Status
Byte register. See Status Reporting for more details concerning
the Operation Status Enable register.
Command Syntax: STATus:OPERation:ENABle <NRf>
Parameter: 0?255
Power-on Value: refers to *PSC command
Example: STATus:OPERation:ENABle 128
Query Syntax: STATus:OPERation:ENABle?
Returned Parameter: <NR1>
Related Commands: *PSC