GMC-I Messtechnik GmbH
5
*SAV
This command saves the parameters of the load in the nonvolatile
memory. There are 10 storage space in total (locations 0 through
9).
Command Syntax: *SAV <NR1>
Parameter: 0~9
Example: *SAV 5
Related Commands: *RCL *RST
*SRE
This command sets the condition of the Status Byte Enable regis-
ter. The programming parameters determine whether or not a 0 to
1 transition in the Status Byte register will set the RQS bit of the
Status Byte register. See
Status Reporting
for more details concern-
ing the Status Byte register.
Command Syntax: *SRE <NR1>
Parameter: 0~255
Query Syntax: *SRE?
Returned Parameter: <NR1>
Related Commands: *PSC
*STB?
This query reads the Status Byte register. The Status Byte register
is cleared when this query has been executed. The *STB? query
and a serial poll share the same returned parameter, but the ser-
vice request bit (bit 6) of the Status Byte register will not be
cleared when a serial poll is sent. See
Status Reporting
for more
details concerning the Status Byte register.
Query Syntax: *STB?
Parameter: None
Returned Parameter: <NR1>
*TRG
This command which is essentially the same as the Group Exe-
cute Trigger (<GET>), generates a trigger to the load only if
TRIG:SOUR is set to BUS. This command becomes effective only
when the trigger source is set to BUS.
Command Syntax: *TRG
Parameter: None
Related Commands: TRIG TRIG:SOUR
*TST?
This query causes the load to go through a limited self-test. The
testing does not alter the mode or parameter settings of the load.
Command Syntax: *TST?
Returned Parameter: <NR1> 0 = test passed None0= a self-
test failure
*WAI
This command instructs the load not process any further com-
mands until all pending operations are completed. All pending
operations are complete including all pending trigger levels have
triggered.
*WAI
can be aborted only by sending the load a GPIB
DCL
(Device Clear) command.
Parameter: None
Related Commands: *OPC *OPC?