[SOURce:]CURRent:[LEVel]:LOW
Specifies the transient current low level, which shares the same
range with the immediate current level.
Command Syntax: [SOURce:]CURRent:[LEVel]:LOWl <NRf+>
Parameter: Digits |MIN|MAX
Unit: A|mA
Example: CURR:LOW 3
Query Syntax: [SOURce:]CURRent:[LEVel]:LOW?
Parameter: None |MIN|MAX
Example: CURR:[LEVel]:LOW? CURR:[LEVel]:LOW? MIN
CURR:[LEVel]:LOW? MAX
Returned Parameter: <NR3>
Related Commands: CURR:LOW
[SOURce:]CURRent:
[LEVel]:HIGH
Specifies the transient current high level, which shares the same
range with the immediate current level.
Command Syntax: [SOURce:]CURRent:[LEVel]:HIGHl <NRf+>
Parameter: Digits |MIN|MAX
Unit: A|mA
Example: CURR:HIGH 3
Query Syntax: [SOURce:]CURRent:[LEVel]:HIGH?
Parameter: None |MIN|MAX
Example: CURR:[LEVel]:HIGH? CURR:[LEVel]:HIGH? MIN
CURR:[LEVel]:HIGH? MAX
Returned Parameter: <NR3>
Related Commands: CURR:HIGH
[SOURce:]CURRent:RISE:RATE
Specifies the current rise rate in CC mode. CURR:RISE:RATE?
MIN and CURR:RISE:RATE? MAX return the maximum and mini-
mum rise rate for the present range.
Command Syntax: [SOURce:]CURRent:RISE:RATE <NRf+>
Parameter: Digits |MIN|MAX
Unit: A|mA/us
Example: CURRent:RISE:RATE 3A
Query Syntax: [SOURce:]CURRent:RISE:RATE?
Parameter: None |MIN|MAX
Example: CURR:RISE:RATE? CURR:RISE:RATE? MIN
CURR:RISE:RATE? MAX
Returned Parameter: <NR3>
Related Commands: CURR:FALL:RATE
[SOURce:]CURRent:FALL:RATE
Specifies the current fall rate in CC mode, which shares the same
range with the current rise rate.
Command Syntax: [SOURce:]CURRent:FALL:RATE <NRf+>
Parameter: Digits |MIN|MAX
Unit: A|mA/us
Example: CURR:FALL:RATE 3A
Query Syntax: [SOURce:]CURRent:FALL:RATE?
Parameter: None |MIN|MAX
Example: CURR:FALL:RATE? CURR:FALL:RATE? MIN
CURR:FALL:RATE? MAX
Returned Parameter: <NR3>
Related Commands: CURR:RISE:RATE
2.2.3
VOLTage Subsystem
[SOURce:]VOLTage[:LEVeI][:IMMediate][:AMPLitude]
Specifies the immediate voltage level for the load. VOLT? MIN and
VOLT? MAX return the maximum and minimum voltage level for
the present range. If the set level exceeds the range, an error will
be generated.
Command Syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<NRf+>
Parameter: Digits |MIN|MAX
Unit: V|mV
Example: VOLT 5V
Query Syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None |MIN|MAX
Example: VOLT? VOLT? MIN VOLT? MAX
Returned Parameter: <NR3>
Related Commands: CURRent, RESistance
[SOURce:]VOLTage[:LEVeI]:TRIGgered[:AMPLitude]
Specifies the voltage trigger level for the load. It shares the same
range with the immediate voltage. The voltage trigger level
becomes the immediate voltage level only when the trigger signal
has been received.
Command Syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
<NRf+>
Parameter: Digits |MIN|MAX
Unit: V|mV
Example: VOLT:TRIG 5V
Query Syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Parameter: None |MIN|MAX
Example: VOLT:TRIG? VOLT:TRIG? MIN VOLT:TRIG? MAX
Returned Parameter: <NR3>
Related Commands: INIT, INIT:CONT, TRIG
[SOURce:]VOLTage:[LEVel]:LOW
Specifies the transient voltage low level, which shares the same
range with the immediate voltage level.
Command Syntax: [SOURce:]VOLTage:[LEVel]:LOW <NRf+>
Parameter: Digits |MIN|MAX
Unit: V|mV
Example: VOLT:LOW 5V
Query Syntax: [SOURce:]VOLTage:[LEVel]:LOW?
Parameter: None |MIN|MAX
Example: VOLT::LOW? VOLT::LOW? MAX VOLT::LOW? MIN
Returned Parameter: <NR3>
Related Commands: VOLT:HIGH, TRAN:LTIM
[SOURce:]VOLTage:[LEVel]:HIGH
Specifies the transient voltage high level, which shares the same
range with the immediate voltage level.
Command Syntax: [SOURce:]VOLTage:[LEVel]:HIGH <NRf+>
Parameter: Digits |MIN|MAX
Unit: V|mV
Example: VOLT:HIGH 5V
Query Syntax: [SOURce:]VOLTage:[LEVel]:HIGHl?
Parameter: None |MIN|MAX
Example: VOLT:HIGH? VOLT:HIGH? MIN VOLT:[LEVel]:HIGH? MAX
Returned Parameter: <NR3>
Related Commands: VOLT:LOW, TRAN:HTIM