Tasks
44
SafeGait 360° Balance and Mobility Trainer |
Instructions for Use | SG360-IFU-004 Rev B
9. Horizontal Lock
HorIzoNTal lock
can be toggled to prevent the actuator from moving horizontally at any point
during a task. Tap the
HorIzoNTal lock
bar, located at the bottom-right of the
Task MoNI-
TorING
screen, to the
oN
or
off
setting (OFF by default). This can otherwise be toggled on the
handheld device by tapping the lock icon on the top-right corner of the screen.
NoTIce:
Do not use HORIZONTAL LOCK for resistance as it may damage the System. If
the patient exerts a large horizontal force on the system, the actuator will move slightly
to protect the patient and the device.
10. RPE Entry
To enable or disable the entry of the patient’s rating of perceived exertion (RPE) upon completion
of the task, slide the
rPe eNTry
bar, located at the bottom right of the
Task MoNITorING
screen to the
oN
or
off
setting, respectively.
For more information about
rPe
, refer to Step 12.
Note: dfP seNsITIVITy, desceNT lIMIT, HorIzoNTal lock,
and
rPe eNTry
can also be set
on the
Task settings
screen by touching
seTTINGs
, located at the bottom right of the
Task
MoNITorING
screen.
NoTIce:
When using the
Task settings
screen on the handheld remote control, take
extra care not to inadvertently make an incorrect selection; for example, touching the
RPE ENTRY On/Off
radio buttons when reaching for the
Descent Limit On/Off
radio
buttons.