GeNeral WarNINGs aNd PrecaUTIoNs
9
Contact Gorbel Medical
| Phone: 1-844-846-8744 | E-mail: [email protected]
•
Connect the system to a main power system (general-purpose alternating-current (AC) electric
power supply) with an earthing or grounding system that provides a protective earth (PE) connection
to avoid the risk of electrical shock.
•
Do not modify the system; in addition to the risks listed above, doing so voids the warranty.
•
Do not attempt to service or repair the system.
•
Do not use the system if maintenance is required.
•
Do not use the system if the actuator is wet; for example, from a sprinkler system.
•
If there is suspicion of electromagnetic interference from other electrical devices, shut down nearby
devices, one-by-one, to locate the source of the interference
•
Ensure that the patient harness connected to the spreader bar is neither too narrow nor too wide;
the performance of the load sensors in the actuator might be adversely affected.
•
Follow the IFU carefully to prevent movement of the patient with sharp, quick motions.
•
Do not manually move the actuator along the track by pulling on the actuator strap.
•
Do not perform jumping-based tasks.
◉
Patients might ascend beyond the upper limit of the device and might adversely affect system
functionality.
◉
This can cause the system to detect a fall due to the accelerated speed of the patient’s down-
ward movement.
• Do not use accessories that are not specified by the manufacturer.
•
Therapists using the system should always practice standard guidelines for Environmental Infection
Control.
• Using a non-SafeGait spreader bar can result in injury.
•
If there is a loss of communication between kiosk and actuator, this will be indicated by a ‘WHITE
LED’ on the status bar and an error message on the kiosk/handheld. If communication loss occurs
during a patient fall it could result in severe discomfort. If communication is not re-established
within two (2) minutes, then follow these steps: 1) Support the patient with assistance from other
staff if needed or other available equipment nearby. 2) Use the Positioning screen to introduce
‘slack’ on the Actuator Strap and Shoulder Straps while continuing to support patient. 3) Disconnect
Shoulder Straps from the Harness while patient is supported. 4) Get patient into safe position such
as chair or mat table. 5) Call Gorbel Medical support.
•
Inspect the harness prior to each use. Do not use if the harness straps are torn or show signs of
wear, there is separation in the harness stitching, or the VELCRO® Brand attachments are unable
to adhere. Leg Cuffs must be attached to the harness torso during use. Contact Gorbel Medical at
1-844-846-8744 to order a replacement harness.
• Positioning the patient outside of the area defined in the Patient Movement Table can result in
premature wear on the strap, increasing the likelihood of injury.
•
While positioning the actuator, ensure the patient is not in the path between current and desired
starting location of the actuator.
•
Do not detach shoulder straps from the spreader bar while patient is being supported. This can lead
to patient injury.
• If the therapist observes that the applied BWS is noticeably more or less than the desired BWS the
patient may become understandably fatigued. In this situation, contact Gorbel Medical for assis
-
tance.