background image

14

15

(correct)

(incorrect )

Blood Glucose Test Strip

Aperture: apply 
the blood to the 
front of the strip.

Black Electrode End: 
Insert this end of test 
strip into meter.

Blood Reaction 
Zone: It must be 
filled with blood 
entirely.

GoodLife System measures the amount of 
glucose in capillary whole blood. Blood can be 
applied to the front of the test strip and is auto-
matically drawn to the test strip through capillary 
action.
GoodLife Blood Glucose Test Strips (Cloudia) are 
intend ed for in vitro diagnostic use with capillary 
whole blood or GoodLife Glucose  Control 

1.

2.

Solution.
Results will not be accurate if used with plasma or 
serum samples.     
Do not use test strips beyond the expiration date 
indicated on the strip vial label.
The discard date for test strips is 90 days after first 
opening the vial. Record the discard date on the 
vial when you open a new vial of test strips.
Blood Glucose Test Strip can be damaged by heat 
and light. Keep them sealed in the original vial. 
Store the vial in a cool, dry place between 
104°F(40°C) and 50°F (10°C). Do not refrigerate.
Do not use damaged test strips in any way. Use 
test strip immediately after taking it out from the 
vial or foil packet; replace the vial cap and close it 
tightly.        

3.

4.

5.

6.

7.

Содержание Cloudia

Страница 1: ...User Guide Blood Glucose Monitoring System Cloudia...

Страница 2: ...Data Transmission 33 Memory Recall 35 Replacing the battery 36 How to Clean and Disinfect 38 Storage and Handling 46 Display Messages and Problem Solving Guide 47 Performance Evaluations System Specif...

Страница 3: ...plays the result as a blood glucose level The GoodLife Blood Glucose Monitoring System Cloudia GoodLife System is designed to provide blood glucose measurements with easy and comfort able testing The...

Страница 4: ...ount of reducing substances such as triglycerides 1000mg dL ascorbic acid 2 25mg dL uric acid 15mg dL and xylose 50mg dL 1 2 3 4 A red blood cell count hematocrit that is either very high over 55 or v...

Страница 5: ...contact at 1 855 692 3511 Mon Fri 9 00 am 4 30 pm Pacific Time NOTE Please make sure that all products listed on the contents of the box are contained in the package before using this system If you f...

Страница 6: ...For Self testing Battery AAA 6V HMD BioMedical Inc Hsinchu Taiwan serial no XXXXX 0000001 SIM 8901204043780370392 For In Vitro Diagnostic Use Only FCC ID AYY0000001 3L 02 XXXXXX V1 SIM AAA AAA AAA AA...

Страница 7: ...be adjusted and stored into 3 types of test modes Before Meal Test After Meal Test and Control Solution Test Alarm Symbol Appears when alarm is set Symbol Appears when data transmission Test Result Ti...

Страница 8: ...lary whole blood or GoodLife Glucose Control 1 2 Solution Results will not be accurate if used with plasma or serum samples Do not use test strips beyond the expiration date indicated on the strip via...

Страница 9: ...check if the GoodLife Blood Glucose Monitoring System Cloudia and Test Strip are working correctly as a system Users should perform a control solution test when a new vial of test strip is opened a m...

Страница 10: ...losed at temperatures between 50 F 10 C and 86 F 30 C Please carefully read the label before use For purchase of control solution please contact your local distributor or customer support at 1 855 692...

Страница 11: ...strip to exit Meter will auto matically turn off NOTE If you do not get OK reading but appear other error message turn off the meter by remove check strip from the meter Then check the battery and rep...

Страница 12: ...below to set the date and time manually Celsius and Fahrenheit can be set according to your preference Then press the Set button to choose between C for Celsius and F for Fahrenheit When Temperature U...

Страница 13: ...ter will display OK before turning off 7 8 Press and hold the Set button for 2 seconds to turn on the meter when ON show up it means auto transmission Press the Set button to choose between ON auto an...

Страница 14: ...sure to tightly replace vial cap after removing test strips Insert test strip immedi ately into strip slot as illustrated The meter turns on automatically Check if the code number on the meter matche...

Страница 15: ...in individuals experiencing a hyperglyce mic hyperosmolar state with or without ketosis and criti cally ill patients should not be tested with the meter Important Safety Information The meter is for...

Страница 16: ...glucose result seems unusually high or low or inconsistent with your previous results check the following 1 Does the code number on the test strip vial match the code number on the meter 2 p i r t s t...

Страница 17: ...tion Setting Auto Data Transmis sion the test result will be transmitted to the database through GSM system after every time you test When the meter is set as manual transmission press es and holds th...

Страница 18: ...g devices with micro processor control such as infusion pumps hemodialysis machines breath ing assistor anesthesia machines infant incubators etc t t t t t t The test records in the meter will not be...

Страница 19: ...t replace the batter ies immediately Remove the old batteries and insert 4 new AAA batteries into the battery compartment Be sure to align the plus and minus signs correctly 4 Close the battery compar...

Страница 20: ...ser To mitigate the risk of bloodborne pathogen trans mission eg viral hepatitis the meter must be prop erly cleaned and disinfected The life of the meter is 5 years for 18 250 cleaning and disinfecti...

Страница 21: ...f premoistened wipe out of canister Remember to recap the canister 4 Wipe entire device until visibly clean Set Mem 5 Throw away the used wipe Please do not reuse the wipe CAUTION 1 The meter should a...

Страница 22: ...d into the test strip slot 5 Throw away the used wipes Please do not reuse the wipe Warning 1 Please use disinfectant with caution 2 For your own safety please do not mix disinfectant with other liqui...

Страница 23: ...infection or need more information please contact your distributor or customer support at 1 855 692 3511 References 1 FDA Public Health Notification Use of Fingerstick Devices on More than one Person...

Страница 24: ...ng battery Please follow all the instructions on the User Guide while operating GoodLife Blood Glucose Meter HMD BioMedical Inc will not be responsible for any impairment occurred from NOT following t...

Страница 25: ...ill ap pears call your doctor immediately Tempera ture is above or below the operating range of test strips The result you have obtained may not be accurate Move to an area with temperature between 50...

Страница 26: ...cient 2 Meter is defective You have to 1 Repeat test with suf ficient sample 2 Perform Meter Check by inserting check strip Error message Err1 Data communi cation module is malfunction Err2 Failure to...

Страница 27: ...g Temp Range 50 F 104 F 10 C 40 C 8 Operating Relative Humidity 20 80 RH 9 Memory Capacity 999 test results with time and date 10 Result Setting Before meal After meal and Control solution test settin...

Страница 28: ...ncentration 75 mg dL Within 5 Within 10 Within 15 Within 20 35 87 40 67 87 77 78 87 90 85 87 98 Tests were conducted by trained technicians in the laboratory setting The venous whole blood from one su...

Страница 29: ...ides an integrated table including test results of blood glucose measurements and quality controls The system does not recommend any medical treat ment or dosing recommendations Hardware requirement P...

Страница 30: ...window on your computer and go to www iglucose net 1 Click on the login button from the iglucose homepage 2 Enter the meter serial number in Email or Serial and Password and click the login button 3 I...

Страница 31: ...60 61 The welcome page will appear Click on the View Readings button 4 Your Time zone page will appear choose your time zone and click on Update Time zone button 5...

Страница 32: ...any of the following tabular or graphic report formats to view your reading Trend Report Pie Chart logbooks or Summary 6 Trend Report Displays high and low glucose read ings within a customized date a...

Страница 33: ...nd post meal glucose readings that fall in and out of target ranges logbooks Displays daily and historical glucose readings with date and time stamp Glucose read ings are also displayed before and aft...

Страница 34: ...ion reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures t 3FPSJFOU PS SFMPDBUF UIF...

Страница 35: ...irement The FCC has granted an Equipment Authorization for this device with all reported SAR levels evaluated as in compliance with the FCCRF exposure guidelines SAR information on this device is on f...

Страница 36: ...ree of defects in material and workmanship for 5 years from the date of original purchase During the stated 5 year period HMD shall at no charge replace a unit found to be defective with an equivalent...

Отзывы: