Disinfection Steps
1. Prepare wipes and meter.
Set
Mem
2. Take a piece of moistened
wipe out of canister.
Remember to recap the
canister.
3. Slowly wipe the meter back and
forth for at least 3 times.
Repeat the same wiping
procedure to all meter
outer surfaces.
4. Make sure the meter stays wet for 1 min.
Please do not get disinfectant liquid
into the test strip slot.
5. Throw away the used wipes.
Please do not reuse the wipe.
Warning
1. Please use disinfectant with caution.
2. For your own safety, please do not mix
disinfectant with other liquid.
3. Please read the instructions carefully on
the canister before starting cleaning and
disinfection.
4. Please don’t use the wipes on any human part.
5. Store disinfectants out of the reach of
children.
6. If any signs of deterioration are noted as below:
Cloudy Display, material cracking, brittle,
softening or meter buttons malfunction,