The GoodLife System is intended for in vitro diagnos-
tic use with capillary whole blood. The system should
not be used for diagnosis and screening of diabetes
or for testing newborn infant (neonatal testing).
CAUTION
The test result does not suggest any medical
indication. The user should consult his or her
clinician for further diagnostic and treatment.
Call your doctor immediately if you experience
symptoms that are not consistent with your
blood glucose test results.
Severe dehydration or excessive water loss may
cause false, high results. Call your doctor right
away if you suspect you are suffering from dehy-
dration.
A sample with large amount of reducing
substances such as triglycerides(>1000mg/dL),
ascorbic acid (>2.25mg/dL), uric acid (>15mg/dL)
and xylose (> 50mg/dL).
1.
2.
3.
4.
A red blood cell count (hematocrit) that is either
very high (over 55%) or very low (under 30%) can
cause false result.
High altitudes above than 8,800 feet (2,750
meter) may affect the test results.
Temperatures outside the range of 50°F to104°F
(10°C to 40°C), 20%-80% RH may affect the test
results.
Do not use GoodLife system to test on critically ill
patient.
Not for use on patients who are dehydrated,
hypotensive,in shock, or in a hyperosmolar state.
5.
6.
7.
8.
9.
6
7
Important Information
Triglyserides
Ascorbic acid
Substance
Highest c oncentration
without interference
( mg/dL)
Toxic
Concertration
( mg/dL)
Theraprutic
Physiological
Levels
N/A
N/A
N/A
N/A
0.40 - 2.00
150 - 500
2.52 - 8.00
57
Uric acid
Xylose
2.25
1000
15
50
Theraprutic/