Internal cleaning (chamber and stainless steel trays sterilization)
NOTE:
Changing in visual appearance inner the autoclave tank is a normal feature of the product
due to the heating of it.
Spots and oxidations signs may occur due to the continuous use and it is totally or partially removed
according to the recommended procedure as below:
In order to execute the internal cleaning of your equipment, sterilization chamber and trays,we
recommend using the product “SDS - 1000”. Product notified at ANVISA/MS.
Follow manufacturer’s directions for cleaning.
Product’s composition: inorganic-acids, fluoride and anionic tense-active based products.
WARNING:
After the cleaning procedure perform a complete cycle of autoclaving and then eliminate
the water of the reservoir and dry the tub using a clean cloth.
NOTE:
The registration at the Ministry of Health of “SDS - 1000” is performed separately from the
product described in this manual, as the “SDS - 1000” is not manufactured by Gnatus.
Internal cleaning (aluminum sterilization chamber)
We recommend cleaning your autoclave monthly with a descaling product (HIGINDOOR 207 Reg.
# MS: 310030079) which cleans the chamber. Apply a solution diluted at a 1:5 (1 part of the product to 5
parts of water) of HIGINDOOR 207 in the area to be cleaned, leaving it to act for 5 minutes, and drying
it with a moist cloth. (check the instructions of use of the product with the manufacturer).
NOTE:
The registration at the Ministry of Health of “HIGINDOOR 207” is performed separately from
the product described in this manual, as the “IHIGINDOOR 207” is not manufactured by Gnatus.
Piping cleaning
Monthly the internal components of the autoclave must be cleaned and disinfected.
To clean the interior of your equipment, tubing, valves and filters, we recommend the use of “INOX
CLEAN” nº Reg. MS: 317130011
Follow manufacturer’s directions for mixing the product.
Product’s composition: inorganic-acids, phosphate, acrilato de sódio. Mix the product following the
instructions given by the manufacturer.
Pour the solution in the chamber and start the sterilizing cycle. Repeat the sterilizing operation using
only distilled water. When these cycles are finished, remove the remnants of the cleaning solution using
a clean piece of cloth moistened with water.
NOTE:
The registration at the Ministry of Health of “INOX CLEAN” is performed separately from the
product described in this manual, as the “INOX CLEAN” is not manufactured by Gnatus.
Cleaning the filter
- The autoclave’s filter (17A), (see page 06) shall be cleaned daily.
- In order to clean the filter, remove it via the handle (17 B), and then wash it in running water.
- Don’t use sharp objects, which could damage the fabric.
Wrapping types for sterilization by autoclave
The materials used for wrapping the pieces to be sterilized must keep the sterilization after the
sterilization process and during storage.
These materials must:
- allow the sterilization process;
- be appropriate for the material and the sterilization process;
- allow the drying, ventilation and dispersion of the sterilizing agents;
- not have lighter and darker parts;
- be resistant to wetness, be free of holes and scratches;
- don’t release particles.
NOTE: GNATUS recommend
the use of surgical paper, because the microbial barrier of this
material is above 90%, and it permits the operator to visually check the state of the instruments. It also
has chemical indicators of the process.
MAINTENANCE