OPERATION
How to use de autoclave
Raise the door latch, push it to the left, and
check the cleanliness of the autoclave.
Connect the energy plug (06) to the socket,
the green led (4D) in the front panel will come on,
indicating that the autoclave is on.
Using the graduated glass (21), pour
inside the autoclave the following quantities
of water:
Autoclave partially full – 150ml
Autoclave completely full – 200ml
WARNING:
Use only distilled water. If you use other water
than distilled, it may cause the obstruction of the
internal valves and stain the instruments, and the
warranty will not be applicable.
All material being sterilized must be previously
washed, rinsed and dried.
IMPORTANT
:
Don’t use “Kraft / TNT” paper
for wrapping the materials to be sterilized (types
of wrappings for autoclave sterilization)
* Observe each country’s standards for
preparation of material to be sterilized.
Distribute the wrappings at equal distances
(22) leaving enough space between them so as
to allow the steam to circulate.
WARNING:
The material to be sterilized (22) can’t be
in contact with any of the internal walls of the
chamber.
Close the autoclave’s door pushing it against
the tank, pushing the door latch to the right and
then lowering it. Check that it is correctly closed
(23)
WARNING:
In order to avoid accidents and burns, it’s very
important that the door is well closed.