Operating manual for the TG uni 1, TG euro 1 and TG euro 1 med appliance tester
Conducting tests in accordance with VDE 0751-1/EN 62353, Example 6 - Testing removable cables
66
Fig. 127
17.6 Documentation
After pressing the [OK] button, the display changes to the „Test result“
screen (Fig. 125).
Here you can see all the measurement results with the related limit values.
If the measurement results of the electrical values, the visual inspection and
the function test were all in order, the following message appears:
„Test OK“.
If the test was not successful the message says:
„Test not OK!“.
At this point you can cancel the test with the [cancel] button (the display
goes back to the „Appliance testing“ screen) or switch to the „Memory
menu“ with the [next] button.
In the „Memory menu“ (Fig. 126) you can confirm or modify the following
parameters: device ID number, device designation, name of the inspector,
date and customer number.
You can also change the storage location – from internal memory to MMC/
SD card and vice-versa.
The current setting is displayed via a symbol in the headline of the memory
menu.
18.
Conducting tests with the appliance tester according
to DIN VDE 0751-1/EN 62353, Example 6: Testing removable
cables
The tests must be performed by qualified personnel. The
qualifications must include technical training, knowledge and
experience of the applicable standards and local regulations.
Personnel assessing the safety of devices must be able to
recognise the possible consequences and dangers resulting from
devices that do not meet the specified requirements. The testing of
the devices must be carried out in compliance with the
accompanying documents.
Connecting the test object
– Connect one end of the test line to the probe socket of the
appliance tester.
– Use the crocodile clip to connect the other end of the test line with
the end of the protective conductor of the cable to be tested.
Ensure that the crocodile clip makes a good contact with the
protective conductor.
– Connect the mains plug of the cable to be tested to the test socket
of the appliance tester.
Fig. 125
Fig. 126
f