12
3. WARNINGS
PERSONNEL SAFETY
The Model 5203 has a low level of fringe field. In operation the magnet
fringing field can be in excess of 0.5mT (5G). This can cause malfunctioning
of heart pacemakers and other medical implants. We recommend that the
fringing field should be mapped and warning signs be placed outside the
0.5mT (5G) contour. Entry to this region of higher field should be restricted
to qualified personnel.
FERROMAGNETIC OBJECTS
During operation the magnet exerts strong magnetic attraction towards
ferromagnetic objects in the near vicinity. Loose objects can be accelerated
to sufficient velocity to cause severe personnel injury or damage to the coils.
Keep ferromagnetic tools clear.
ARCING
This magnet stores considerable energy in its field during operation. Do not
disconnect any current lead while under load or the magnetic field energy
will be discharged across the interruption causing hazardous arcing.
HIGH VOLTAGES and CURRENTS
Be aware that the AC mains input to the DC power supply and the output
to the electromagnet are high voltage. Follow lock-out / tag-out procedures
during all maintenance procedures. Verify that all electrical connections
are secure and properly strain relieved. Loose connections may overheat
and cause damage to equipment.
HOT COIL RESISTANCE
Do not exceed the maximum coil hot resistance given in the specifications or
coil overheating and possible damage may occur.
Содержание 5203
Страница 2: ...Blank Page ...
Страница 11: ...9 2 3 11907 0224 0 5203 General Assembly ...
Страница 12: ...10 2 4 11907 0233 0 5203 Terminal Block Assembly ...
Страница 13: ...11 2 5 17907 0412 0 5203 Mounting Plate ...
Страница 24: ...22 9 EXCITATION CURVES AND FIELD UNIFORMITY Excitation Curves for the various pole types ...
Страница 25: ...23 Excitation and Field Remanence for the standard pole Type A ...
Страница 26: ...24 Near and far vertical field profiles for the standard pole Type A ...
Страница 27: ...25 ...
Страница 28: ...26 ...
Страница 39: ...37 10 TEST DATA ...
Страница 41: ...39 13907 0025 1 System Wiring 5203 1x BOP20 50 and PDU ...
Страница 42: ...40 13907 0026 1 System Wiring 5203 2x BOP20 50 and PDU ...
Страница 43: ...41 1C907 5203 08 Electrical Assembly 5203 and IECO BPS ...
Страница 44: ...42 1D907 5203 08 Electrical Wiring 5203 and IECO BPS ...
Страница 45: ...43 1C907 5203 09 Electrical Assembly 5203 IECO BPS and 5972 ...
Страница 46: ...44 1D907 5203 09 Electrical Wiring 5203 IECO BPS and 5972 ...
Страница 48: ...46 16907 0116 1 S1 Cable Current Interlock 5203 BOP20 50 and PDU ...
Страница 49: ...47 16907 0116 3 S1 Cable Current Interlock 5203 IECO BPS and 5972 ...
Страница 50: ...48 16907 0281 0 Cable Assembly Interlock Pigtail 5972 or IECO BPS ...
Страница 51: ...49 11907 0620 0 GMW RC 1920 Equipment Rack ...
Страница 57: ...BPS 40 100 Bipolar Power Supply BPS 40 100 40V 100A International Electric Co Document 200203PHa 3 2 2020 ...
Страница 65: ...ThermoFlex Preventative Maintenance ...