8. Operation    

CAUTION

•  Do not push or lean against the centrifuge while it is running.

•  Do not run the centrifuge when fragments or sample solutions are left in the centrifuge chamber. Always keep 

the centrifugal chamber clean. 

•  If the centrifuge makes a strange noise during operation, stop it immediately and contact Globe Scientific. 

Please include any warning code(s), if displayed.

8.1 Normal Operation

Turn on the power switch and the centrifuge will start self-diagnostic checks, see figure 8-1 below:

After self-diagnostics, the centrifuge will display the cumulative running time, see figure 8-2 below:

Figure 8-2 indicates the centrifuge has accumulated running time 312 hours 56 minutes and 45 seconds.  
Then the centrifuge displays the last running parameters, see figure 8-3 below:

Speed: 15000rpm; running time: 12 minutes; centrifugal chamber temperature: 25ºC. 
The lid lock is then released. 

HELLO

88

88

ºC

min

00312

56

45

15000

25

12

ºC

min

rpm

Figure 8-1 Self-check interface

Figure 8-2 Cumulative running time interface

Figure 8-3 Last running interface

Содержание GCM-24

Страница 1: ...eed MicroCentrifuges USER MANUAL GCM 24 GCM R 24 www globescientific com IMPORTANT Before using this centrifuge please carefully and completely read this user manual for efficient operation and proper...

Страница 2: ...hooting Tips 15 16 10 1 Possible Problems and Solutions 15 10 2 How to Open the Lid 15 10 3 Replacement of Fuses 16 11 Rotor Instructions 16 19 11 1 Rotor Structure 16 12 Calculating RCF 17 13 Returni...

Страница 3: ...FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected incidents B...

Страница 4: ...mmediately and contact our service center Notify the service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are like...

Страница 5: ...otection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in a...

Страница 6: ...g on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Keep obstructi...

Страница 7: ...Lid Seal Ring Lid Seal Ring Control Panel Control Panel Power Switch Power Switch Power Switch Power Switch Lid Lock Hook Lid Lock Hook Cover Air Vents Outlet View Port View Port Air Vents Air Vents...

Страница 8: ...Adjustment Knob Rotate clockwise to increase parameter values Rotate anti clockwise to decrease parameter values Press the knob to switch between Speed and RCF 1 2 3 4 5 The main display is shown abo...

Страница 9: ...e the parameter you wish to modify Adjustment Knob Rotate clockwise to increase parameter values Rotate anti clockwise to decrease parameter values Press the knob to switch between Speed and RCF 1 2 3...

Страница 10: ...be used with the centrifuge 7 2 Symmetrically Load Tubes Into Rotor IMPORTANT The rotor should always be loaded in a balanced manner to avoid personal injury and to ensure centrifuge longevity When an...

Страница 11: ...and the centrifuge will start self diagnostic checks see figure 8 1 below After self diagnostics the centrifuge will display the cumulative running time see figure 8 2 below Figure 8 2 indicates the c...

Страница 12: ...ighten the rotor to the shaft Close the rotor lid firmly tighten the lid clockwise to the rotor and ensure it is positioned properly Close the door and then start running The method of removing the ro...

Страница 13: ...me from 30 seconds up to 99 minutes When the time displays HD the centrifuge is in continuous running mode 3 Set the temperature GCM R 24 Press the Select button the temperature value will flash into...

Страница 14: ...rs section 11 Press the Select button and choose speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the...

Страница 15: ...gs If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge by wiping with a cloth moistened with a 70 ethanol solution 3 Dri...

Страница 16: ...is a short circuit between the power cord grounding pin and the motor shaft If an open circuit is found contact Globe Scientific before use Replaceable Wearing Parts Replacement conditions Lid Sealin...

Страница 17: ...ring operation Close the door immediately Press the Start Stop button while the door is opening Close the door and then start normal operation E 04 Temperature fault Air vents are blocked Clean air ve...

Страница 18: ...orward to release the lid 10 3 Replacing Fuses 1 There are two fuses in the GCM 24 250V 5A time delay type size 5 20 There are two fuses of GCM R 24 250V 10A time delay type size 5 20 2 The fuse holde...

Страница 19: ...mum speed rpm Maximum centrifugal radius cm Maximum RCF g GCM 24 2ML 2 0 1 5ml tube 15000 8 5 21380 0 2ml PCR tube GCM 24 AD2 15000 6 9 17350 0 5ml PCR tube GCM 24 AD5 15000 7 6 19100 GCM 24 05ML 0 5m...

Страница 20: ...of the individual EC countries may vary If necessary contact your supplier 14 Ordering Information Item Descriptions GCM 24 Centrifuge Micro 24 Place High Speed 120v 60Hz US Plug Includes 24 Place Rot...

Страница 21: ...use with GCM 24 Series High Speed Micro Centrifuges Aluminum Alloy Fixing Clips and O Ring 15000rpm max 36 Place for 0 5mL Microcentrifuge Tubes GCM 24 PCR Rotor with Lid for use with GCM 24 Series Hi...

Страница 22: ...roperly Instances that would void the warranty include 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by transportation or relocation aft...

Отзывы: