8. Operation (Cont.)

START THE OPERATION

Press the Select button to select the desired parameters. The parameters can be modified when the value is flashing. 
Rotate the Adjustment Knob clockwise to increase parameter value. Rotate the Adjustment Knob counter clockwise to 
decrease parameter value. Rotate the Adjustment Knob faster and the parameter value will increase faster. The minimum 
speed increment is 10 rpm, the minimum time increment is 1 second.
1. Set the speed

• Press the Select button until the speed rpm is displayed.

• When the Speed button is selected, the speed symbol will flash the speed value.

• The minimum speed value you can set is 500rpm; the minimum increment is 100 rpm.

• Rotate the Adjustment Knob clockwise to increase speed value. Rotate the Adjustment Knob anti-clockwise to 

decrease speed value.

• There is a circulating function to increase/decrease the speed values. Rotate the Adjustment Knob clockwise to 

change settings from small » large » maximum » minimum. Rotate the Adjustment Knob anti-clockwise to change 
settings from large » small » minimum » maximum.

2. Set the time

• Press the Select button, the time value flashes in the time setting mode.

• Rotate the Adjustment Knob to set the running time from 30 seconds up to 99 minutes.

• When the time displays HD, the centrifuge is in continuous running mode.

3. Set the temperature (GCM-R-24)

• Press the Select button, the temperature value will flash into temperature set mode.

• Rotate the Adjustment Knob, and input a temperature value ranging from -20ºC to 40ºC.

• When the Temperature Symbol   flashes, this indicates the refrigeration system is running, not flashing indicates the  

refrigeration is not running.

4. Press the 

Start / Stop

 button to start running

• The lid must be locked before rotor starts rotating.

• The timer will begin counting down once the speed setting value is reached, the screen then displays the  

 

remaining run time.

5. View and modify the operation programs

• Pressing the Select button returns the display to the programming interface and displays program setting. Press the 

Select button to the desired program. When flashing, rotate the Adjustment Knob to modify values. Release the 
button after 5 seconds, and the centrifuge will return to normal operation mode and run according to the new value.

• If the set time value has been modified, the operation time is not affected and will continue.

6. Warning display

• If an error occurs during operation the centrifuge will brake to stop automatically and then display the error code on the 

time/display area. The error code can be checked in table 10-1, and corrective actions can be performed accordingly.

Содержание GCM-24

Страница 1: ...eed MicroCentrifuges USER MANUAL GCM 24 GCM R 24 www globescientific com IMPORTANT Before using this centrifuge please carefully and completely read this user manual for efficient operation and proper...

Страница 2: ...hooting Tips 15 16 10 1 Possible Problems and Solutions 15 10 2 How to Open the Lid 15 10 3 Replacement of Fuses 16 11 Rotor Instructions 16 19 11 1 Rotor Structure 16 12 Calculating RCF 17 13 Returni...

Страница 3: ...FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected incidents B...

Страница 4: ...mmediately and contact our service center Notify the service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are like...

Страница 5: ...otection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in a...

Страница 6: ...g on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Keep obstructi...

Страница 7: ...Lid Seal Ring Lid Seal Ring Control Panel Control Panel Power Switch Power Switch Power Switch Power Switch Lid Lock Hook Lid Lock Hook Cover Air Vents Outlet View Port View Port Air Vents Air Vents...

Страница 8: ...Adjustment Knob Rotate clockwise to increase parameter values Rotate anti clockwise to decrease parameter values Press the knob to switch between Speed and RCF 1 2 3 4 5 The main display is shown abo...

Страница 9: ...e the parameter you wish to modify Adjustment Knob Rotate clockwise to increase parameter values Rotate anti clockwise to decrease parameter values Press the knob to switch between Speed and RCF 1 2 3...

Страница 10: ...be used with the centrifuge 7 2 Symmetrically Load Tubes Into Rotor IMPORTANT The rotor should always be loaded in a balanced manner to avoid personal injury and to ensure centrifuge longevity When an...

Страница 11: ...and the centrifuge will start self diagnostic checks see figure 8 1 below After self diagnostics the centrifuge will display the cumulative running time see figure 8 2 below Figure 8 2 indicates the c...

Страница 12: ...ighten the rotor to the shaft Close the rotor lid firmly tighten the lid clockwise to the rotor and ensure it is positioned properly Close the door and then start running The method of removing the ro...

Страница 13: ...me from 30 seconds up to 99 minutes When the time displays HD the centrifuge is in continuous running mode 3 Set the temperature GCM R 24 Press the Select button the temperature value will flash into...

Страница 14: ...rs section 11 Press the Select button and choose speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the...

Страница 15: ...gs If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge by wiping with a cloth moistened with a 70 ethanol solution 3 Dri...

Страница 16: ...is a short circuit between the power cord grounding pin and the motor shaft If an open circuit is found contact Globe Scientific before use Replaceable Wearing Parts Replacement conditions Lid Sealin...

Страница 17: ...ring operation Close the door immediately Press the Start Stop button while the door is opening Close the door and then start normal operation E 04 Temperature fault Air vents are blocked Clean air ve...

Страница 18: ...orward to release the lid 10 3 Replacing Fuses 1 There are two fuses in the GCM 24 250V 5A time delay type size 5 20 There are two fuses of GCM R 24 250V 10A time delay type size 5 20 2 The fuse holde...

Страница 19: ...mum speed rpm Maximum centrifugal radius cm Maximum RCF g GCM 24 2ML 2 0 1 5ml tube 15000 8 5 21380 0 2ml PCR tube GCM 24 AD2 15000 6 9 17350 0 5ml PCR tube GCM 24 AD5 15000 7 6 19100 GCM 24 05ML 0 5m...

Страница 20: ...of the individual EC countries may vary If necessary contact your supplier 14 Ordering Information Item Descriptions GCM 24 Centrifuge Micro 24 Place High Speed 120v 60Hz US Plug Includes 24 Place Rot...

Страница 21: ...use with GCM 24 Series High Speed Micro Centrifuges Aluminum Alloy Fixing Clips and O Ring 15000rpm max 36 Place for 0 5mL Microcentrifuge Tubes GCM 24 PCR Rotor with Lid for use with GCM 24 Series Hi...

Страница 22: ...roperly Instances that would void the warranty include 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by transportation or relocation aft...

Отзывы: