This section describes the instructions that you should adhere to when installing the centrifuge to ensure your safety and
optimum performance of the unit. Before moving the centrifuge, the rotor must be removed.
WARNING
• Improper power supply may damage centrifuge
• Make sure the power source conforms to the required power supply before connecting
4.1 Location
1. Place this centrifuge on a firm, flat and level surface and ensure the four feet of this centrifuge stand on the counter
firmly. Avoid installing on a slippery surface or a surface prone to vibration.
2. Ideal ambient temperature is 20ºC ± 5ºC, avoid placing the centrifuge in direct sunlight if the ambient temperature
exceeds 30ºC.
3. Keep obstructions clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to ensure optimal
cooling efficiency.
4. Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
4.2 Connection of the Power Cord and Grounding
WARNING
• To avoid electrical shocks, ensure your hands are dry when touching the power cord.
• This centrifuge must be grounded properly.
A minimum 10A outlet providing a sufficient ground is required, and this must meet local safety requirements.