3
2. Declaration of Conformity
Construction in accordance with the following safety standards:
EN 61010-1
EN 61010-2-020
EN 61010-2-101
Construction in accordance with the following EMC standards:
EN 61326-1/ FCC Part 15 Subpart B/ IECS 001
EN 61326-2-6:2006
Associated EU guidelines:
EMC directive: 2004/108/EC
LVD directive: 2006/95/EC
IVD directive: 98/79/EC
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority
to operate the equipment.
NOTE:
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment
in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his
own expense.
3. Required Operating Conditions
3.1 Basic Operational Conditions
• Power: 110V-240V, 50Hz/60Hz, 3A
• Ambient temperature: 2ºC~40ºC
• Relative humidity: ≤80%, non-condensing
• No vibration and airflow around the centrifuge
• No static, explosive and corrosive gases around the centrifuge
3.2 Transport and Storage Conditions
• Storage temperature: -40ºC-55ºC
• Relative humidity: ≤93%