9. Maintenance                          

9.1 Cleaning  

   

CAUTION

•  Failure to follow the recommended instructions for cleaning and disinfecting may result in damage to the 

centrifuge.

1. Centrifuge

• If the centrifuge is exposed to ultraviolet rays for a long period of time, the lid may discolor or the label may be peel off. 

It is recommended to cover the centrifuge with a piece of cloth to protect it from direct exposure or to move it from 
direct exposure.

• If the centrifuge needs cleaning, clean it with a cloth or sponge moistened with a neutral detergent solution.
• Sterilize the centrifuge by wiping with a cloth moistened with a 70% ethanol solution.

2. Rotor Chamber

  
 CAUTION

•  Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber, otherwise fluids 

may leak into the drive units and cause corrosion or deterioration to the bearings. 

• If the rotor chamber needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution. Sterilize 

the centrifuge by wiping with a cloth moistened with a 70% ethanol solution.

3. Drive Shaft

• We recommend regular maintenance for the drive shaft. You can wipe the drive shaft with a soft cloth, and then apply 

a thin coat of silicon grease.

4. Lid

• Clean and sterilize the lid using the same method as in step one above.

5. Rotor

• To prevent corrosion, remove the rotor from the rotor chamber. If not in use for a long period of time, turn the rotor 

upside down to dry the tube cavities and keep clean.

• For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor when it is 

completely dry.

• The rotor should be checked every 3 months to ensure the tube and rotor cavities are clean and to apply a thin coat of 

silicon grease.

Содержание GCC-E

Страница 1: ...l Centrifuge Enhanced Feature Low Speed Clinical Centrifuge USER MANUAL GCC E www globescientific com IMPORTANT Carefully and completely read this user manual to ensure user safety and achieve most ef...

Страница 2: ...ps 10 11 10 1 How to Open the Lid 11 11 Rotor Instructions 12 13 11 1 Rotor Structure 12 11 2 Available Rotors and Adapters 12 12 Calculating RCF 13 13 Returning and Disposal 13 13 1 Returning the Uni...

Страница 3: ...ASK FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected inciden...

Страница 4: ...he service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are likely to damage the centrifuge Please contact our ser...

Страница 5: ...uant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment gener...

Страница 6: ...d installing on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Kee...

Страница 7: ...s to start the centrifuge Press to stop if centrifuge is already running Adjustment Knob Rotate clockwise to increase parameter values Rotate counterclockwise to decrease parameter values Press to swi...

Страница 8: ...sure centrifuge longevity When any tube is loaded a tube of the same type volume and weight must be added in the opposite rotor location Never intentionally run the centrifuge in an unbalanced conditi...

Страница 9: ...nimum Rotate the Adjustment Knob counter clockwise to change settings from large small minimum maximum 2 Set the time Press the Select button the time value flashes in the time setting mode Rotate the...

Страница 10: ...speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the input mode will be shut down Rotate the Adjustme...

Страница 11: ...may leak into the drive units and cause corrosion or deterioration to the bearings If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize...

Страница 12: ...and the operator can determine the malfunction with the alarm code below 9 Maintenance Cont 9 2 Rotor Installation 1 Installation 2 Adjustment 1 2 3 4 IMPORTANT Before locking the rotor rotate it and...

Страница 13: ...operation E 06 The set parameter exceeds the Modify the set parameter Incorrect allowable range value Parameter E 10 86 Refer to service manual Contact Globe Scientific 10 1 How to Open the Lid WHEN...

Страница 14: ...Dimension Adapters Maximum Radius Maximum Type Type Per Rotor L mm Required Speed rpm cm RCF g 15mL Conical 8 17 120 None 4500 11 2490 1 5 5mL 12 13 82 CC AD1 4500 9 8 2218 Vacuum CC PLUG 4 7mL 13 106...

Страница 15: ...be disposed of as part of domestic waste The device belongs to group 8 medical devices and is categorized in the business to business field The icon of the crossed out trash can shows that the device...

Страница 16: ...on 5 Failures caused by using non standard spare parts or accessories and or unauthorized modification of the rotor or centrifuge 6 Failures caused by natural disasters including fire earthquakes etc...

Отзывы: