Globe Scientific GCC-E Скачать руководство пользователя страница 13

11

• Alarm codes E-1~E-9 are related to incorrect operation/programming. You can continue running the centrifuge after 

implementing corrective procedures.

10. Troubleshooting Tips (Cont.)

Problem 

 

Cause 

Solution

Nothing displays on the 

 

Building circuit breaker or GFI 

Reset the circuit breaker or 

screen when the Power 

 

recepticle has tripped 

GFI recepticle and turn on 

Switch is turned on 

 

 

Power Switch

Abnormal Vibration

 

 

Rotor misaligned with motor shaft 

Reinstall rotor as shown in  

 

 

 

Section 9.2    

  

 

  

Samples are unbalanced 

 

Ensure tube weights match, 

 

 

 

install symmetrically as   

 

 

 

described in section 7.2 

Alarm Code Displayed

 E-02

 

The lid opened during operation 

Close the lid immediately

 

 

Lid Fault

 

 
 

 The 

Start/Stop button was 

Close the lid and then start 

 

 

pressed while the lid was open 

normal operation 

 

E-06 

The set parameter exceeds the 

Modify the set parameter

 

 Incorrect 

allowable range 

value

 

 Parameter 

 
 

E-10~86 

Refer to service manual 

Contact Globe Scientific

10.1 How to Open the Lid

WHEN POWERED ON 

  
 CAUTION

•  The lid should only be opened when the rotor has come to a complete stop.

1. Turn on the power switch; the lid is then released automatically.

2. The lid lock will be released automatically once the operation is finished.

3. It is possible to release the lid by pressing the 

Door

 button when the rotor has stopped.

IN THE CASE OF POWER OUTAGE 

The lid will not open automatically if there is a power outage. It is possible to open manually as follows:

1. Ensure the rotor has stopped rotating.

• Listen carefully to ensure no rotation can be heard.

2. Insert a screw driver into each lid release hole to open the lid.

• Holes are located on the left and right sides of the unit.
• Carefully insert a screw driver into the lid release holes and push forward to release the lid.

Содержание GCC-E

Страница 1: ...l Centrifuge Enhanced Feature Low Speed Clinical Centrifuge USER MANUAL GCC E www globescientific com IMPORTANT Carefully and completely read this user manual to ensure user safety and achieve most ef...

Страница 2: ...ps 10 11 10 1 How to Open the Lid 11 11 Rotor Instructions 12 13 11 1 Rotor Structure 12 11 2 Available Rotors and Adapters 12 12 Calculating RCF 13 13 Returning and Disposal 13 13 1 Returning the Uni...

Страница 3: ...ASK FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected inciden...

Страница 4: ...he service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are likely to damage the centrifuge Please contact our ser...

Страница 5: ...uant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment gener...

Страница 6: ...d installing on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Kee...

Страница 7: ...s to start the centrifuge Press to stop if centrifuge is already running Adjustment Knob Rotate clockwise to increase parameter values Rotate counterclockwise to decrease parameter values Press to swi...

Страница 8: ...sure centrifuge longevity When any tube is loaded a tube of the same type volume and weight must be added in the opposite rotor location Never intentionally run the centrifuge in an unbalanced conditi...

Страница 9: ...nimum Rotate the Adjustment Knob counter clockwise to change settings from large small minimum maximum 2 Set the time Press the Select button the time value flashes in the time setting mode Rotate the...

Страница 10: ...speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the input mode will be shut down Rotate the Adjustme...

Страница 11: ...may leak into the drive units and cause corrosion or deterioration to the bearings If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize...

Страница 12: ...and the operator can determine the malfunction with the alarm code below 9 Maintenance Cont 9 2 Rotor Installation 1 Installation 2 Adjustment 1 2 3 4 IMPORTANT Before locking the rotor rotate it and...

Страница 13: ...operation E 06 The set parameter exceeds the Modify the set parameter Incorrect allowable range value Parameter E 10 86 Refer to service manual Contact Globe Scientific 10 1 How to Open the Lid WHEN...

Страница 14: ...Dimension Adapters Maximum Radius Maximum Type Type Per Rotor L mm Required Speed rpm cm RCF g 15mL Conical 8 17 120 None 4500 11 2490 1 5 5mL 12 13 82 CC AD1 4500 9 8 2218 Vacuum CC PLUG 4 7mL 13 106...

Страница 15: ...be disposed of as part of domestic waste The device belongs to group 8 medical devices and is categorized in the business to business field The icon of the crossed out trash can shows that the device...

Страница 16: ...on 5 Failures caused by using non standard spare parts or accessories and or unauthorized modification of the rotor or centrifuge 6 Failures caused by natural disasters including fire earthquakes etc...

Отзывы: