11
• Alarm codes E-1~E-9 are related to incorrect operation/programming. You can continue running the centrifuge after
implementing corrective procedures.
10. Troubleshooting Tips (Cont.)
Problem
Cause
Solution
Nothing displays on the
Building circuit breaker or GFI
Reset the circuit breaker or
screen when the Power
recepticle has tripped
GFI recepticle and turn on
Switch is turned on
Power Switch
Abnormal Vibration
Rotor misaligned with motor shaft
Reinstall rotor as shown in
Section 9.2
Samples are unbalanced
Ensure tube weights match,
install symmetrically as
described in section 7.2
Alarm Code Displayed
E-02
The lid opened during operation
Close the lid immediately
Lid Fault
The
Start/Stop button was
Close the lid and then start
pressed while the lid was open
normal operation
E-06
The set parameter exceeds the
Modify the set parameter
Incorrect
allowable range
value
Parameter
E-10~86
Refer to service manual
Contact Globe Scientific
10.1 How to Open the Lid
WHEN POWERED ON
CAUTION
• The lid should only be opened when the rotor has come to a complete stop.
1. Turn on the power switch; the lid is then released automatically.
2. The lid lock will be released automatically once the operation is finished.
3. It is possible to release the lid by pressing the
Door
button when the rotor has stopped.
IN THE CASE OF POWER OUTAGE
The lid will not open automatically if there is a power outage. It is possible to open manually as follows:
1. Ensure the rotor has stopped rotating.
• Listen carefully to ensure no rotation can be heard.
2. Insert a screw driver into each lid release hole to open the lid.
• Holes are located on the left and right sides of the unit.
• Carefully insert a screw driver into the lid release holes and push forward to release the lid.