2

  
CAUTION

•  This centrifuge must be located on a firm and level table.

•  Make sure the centrifuge is horizontal before running.

•  Do not move or relocate the centrifuge when it is running.

•  If fluid spills in the rotor chamber, please promptly clean and dry with a dry cloth to avoid sample 

contamination. 

•  Ensure removal of any objects and/or fragments of the tubes dropped inside the rotor chamber before running 

the centrifuge.

Rotor Cautions: 
•  Always check for corrosion and damage on the rotor surface before using it. Do not use the rotor if an 

abnormality is found.

•  Do not set the speed beyond the allowable maximum speed of the rotor kits (rotor and adapters). 

•  Do not exceed the allowable imbalance.

•  Use the rotor and tubes within their actual capacities.

•  If any abnormal condition occurs during operation, please stop immediately and contact our service center. 

Notify the service center of the warning code (if displayed) along with the model and serial numbers and a 
detailed description of the issue. 

•  Vibrations are likely to damage the centrifuge.  Please contact our service center if any abnormality is observed.

1. Specifications                         

Maximum Speed 

4500rpm(300-4500rpm), increment: 100rpm

Maximum RCF 

2490×g, increment: 100×g

Maximum Capacity 

10ml×12, 15ml×8

Timer 

30seconds - 99minutes-HOLD, continuous operation

Noise 56dB(A)
Driving Motor 

Brushless DC motor 

Safety Devices 

Lid interlock, over-speed detector, automatic internal diagnosis

Power Requirements 

Single-phase, 110V-240V, 50Hz/60Hz, 3A. 

Ambient Condition
   -Set-Up Site 

Indoors only

   -Altitude 

Up to 2000 m above sea level

   -Ambient Temperature 

2°C ~ 40°C 

   -Humidity 

80%

   -Excess-Voltage Category 

II

   -Pollution Degree 

Device Protection Class 

I

EMC 

EN/IEC 61326-1 

FCC  Class A

   -Emitted Interference,
  

 -

Interference Immunity 

Class A 

Dimensions (mm) 

280 × 364 × 266 (L x W x H)

Weight 6kg 
Additional features 

Speed/RCF switch, pulse operation, processing display, audible reminder

Important Safety Information (Cont.)

Содержание GCC-E

Страница 1: ...l Centrifuge Enhanced Feature Low Speed Clinical Centrifuge USER MANUAL GCC E www globescientific com IMPORTANT Carefully and completely read this user manual to ensure user safety and achieve most ef...

Страница 2: ...ps 10 11 10 1 How to Open the Lid 11 11 Rotor Instructions 12 13 11 1 Rotor Structure 12 11 2 Available Rotors and Adapters 12 12 Calculating RCF 13 13 Returning and Disposal 13 13 1 Returning the Uni...

Страница 3: ...ASK FOR HELP The precautions described in this User Manual are carefully developed in an attempt to cover all the possible risks However it is also important that you are alert for unexpected inciden...

Страница 4: ...he service center of the warning code if displayed along with the model and serial numbers and a detailed description of the issue Vibrations are likely to damage the centrifuge Please contact our ser...

Страница 5: ...uant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment gener...

Страница 6: ...d installing on a slippery surface or a surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the centrifuge in direct sunlight if the ambient temperature exceeds 30 C 3 Kee...

Страница 7: ...s to start the centrifuge Press to stop if centrifuge is already running Adjustment Knob Rotate clockwise to increase parameter values Rotate counterclockwise to decrease parameter values Press to swi...

Страница 8: ...sure centrifuge longevity When any tube is loaded a tube of the same type volume and weight must be added in the opposite rotor location Never intentionally run the centrifuge in an unbalanced conditi...

Страница 9: ...nimum Rotate the Adjustment Knob counter clockwise to change settings from large small minimum maximum 2 Set the time Press the Select button the time value flashes in the time setting mode Rotate the...

Страница 10: ...speed unit g the speed symbol will then flash into RCF value input status If no button is pressed after the speed value has flashed after 5 seconds the input mode will be shut down Rotate the Adjustme...

Страница 11: ...may leak into the drive units and cause corrosion or deterioration to the bearings If the rotor chamber needs cleaning clean with cloth or sponge moistened with a neutral detergent solution Sterilize...

Страница 12: ...and the operator can determine the malfunction with the alarm code below 9 Maintenance Cont 9 2 Rotor Installation 1 Installation 2 Adjustment 1 2 3 4 IMPORTANT Before locking the rotor rotate it and...

Страница 13: ...operation E 06 The set parameter exceeds the Modify the set parameter Incorrect allowable range value Parameter E 10 86 Refer to service manual Contact Globe Scientific 10 1 How to Open the Lid WHEN...

Страница 14: ...Dimension Adapters Maximum Radius Maximum Type Type Per Rotor L mm Required Speed rpm cm RCF g 15mL Conical 8 17 120 None 4500 11 2490 1 5 5mL 12 13 82 CC AD1 4500 9 8 2218 Vacuum CC PLUG 4 7mL 13 106...

Страница 15: ...be disposed of as part of domestic waste The device belongs to group 8 medical devices and is categorized in the business to business field The icon of the crossed out trash can shows that the device...

Страница 16: ...on 5 Failures caused by using non standard spare parts or accessories and or unauthorized modification of the rotor or centrifuge 6 Failures caused by natural disasters including fire earthquakes etc...