Section 2 General Information
2-4
WARNING
!
DO NOT REMOVE THE STRAP OR 2 X 4 WOOD BRACE HOLDING THE SEAT IN
PLACE UNTIL THE CHAIR IS IN ITS PERMANENT LOCATION.
1. Remove the fasteners along the bottom of the packaging that secure the chair to the wooden pallet.
2.
Lift the chair off of the skid by grasping its back and seat framework. Do not lift by the arms or headrest.
See Figure 2-1
3.
When chair is in its permanent position cut and remove strap.
4. Plug the chair into the wall once the chair is placed in its permanent location.
5. Press the “SEAT UP” button to lift the chair off of the 2 x 4 wood brace. Be sure to lift the chair high enough
so the bottom of the calf rest is above the 2 x 4 wood brace.
6. Slide the 2 x 4 wood brace off the chair base and set aside.
7.
All shipping materials should be retained until it has been determined that the unit was not damaged during
shipment.
If damage is discovered, completed the following:
• Do not refuse shipment.
• Make a notation on the delivery receipt, and inspect the carton for damage.
• Take pictures of damage to the equipment, and to the packaging (if evident).
• If damage is discovered leave in original container and request immediate inspection from the carrier.
•
Contact the Global Surgical Customer Service Department at 1-800-861-3585.
• If the product is damaged electrically or mechanically and in the event the original packing materials are no
longer available, contact Global Surgical
Customer Services Department.
•
Refer to Section 7 of this manual regarding service and warranty and proceed as instructed.