PART TWO - INSTALLATION
SECTION 2 - MECHANICAL REQUIREMENTS
Rev. 2-300
Accu4™ Low-range Turbidimeter System
23
2.4 Analyzer Location
2.5 Analyzer Mounting
1.
Locate the analyzer as close as possible to the installed
sensor. Do not exceed a maximum distance of 30 feet
(9 m) between the sensor and analyzer. (Consult fac-
tory if longer distances are required.)
2.
Mount the analyzer in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-4 to +140
°
F or
-20 to +60
°
C).
CAUTION:
EXPOSING THE ANALYZER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED LIMIT.
Figure 2-6 illustrates various ways to mount the analyzer
using the supplied bracket and hardware. Determine the
mounting method and attach the hardware as shown in the
respective illustration. Refer to Figure 2-7 for analyzer in-
stallation dimension details.
FIGURE 2-6 Analyzer Mounting Arrangements