EMC DECLARATIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The
UTC2
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
UTC2
should assure that is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
UTC2
uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
UTC2
is suitable for use in all establishments, including domestic establishments and those directly con-
nected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The
UTC2
is intended for use in the electromagnetic environment specified below. The customer or the user of the
UTC2
should assure that is used
in such an environment
Immunity test
IEC60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic dis-
charge (ESD)
IEC61000-4-2
contact ± 6 kV
air ± 8 kV
contact± 6 kV
air ± 8 kV
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast tran-
sient/burst
IEC61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 s
< 5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5 s
Mains power quality should be that of a typical commercial or hospital environment.
If the user of
UTC2
requires continued operation during power mains interruptions, it is recommended
that the
UTC2
be powered from a uninterruptible power supply or a battery.
Содержание UTC2
Страница 4: ...MA531_EN EN 2 UTC2...
Страница 9: ...English EN MA531_EN EN 7 GMAC12 00607 01 00607 02 00607 03 00602 040 00602 060 00602 100...
Страница 43: ...English EN MA531_EN EN 41 ET Treatment Scheme Current CH1 CH2 R N R N...
Страница 47: ...English EN MA531_EN EN 45 Combi Treatment Scheme CH1 CH2 R N N...
Страница 56: ...e n nt lt...