acoustic signal.
IMPORTANT:
To avoid the danger of the formation of stationary waves it is necessary to move the handpiece with circular
or coming and going slow and regular movement. In case of treatment with static handpiece, select a low modulation level
(5/10%) and low intensity (0.5-1.0W/cm
2
).
IMPORTANT:
To avoid the damage of the ultrasound transducer, start the power delivery after having the handpiece in
energy dissipative condition.
US Treatment Scheme (Ultrasound)