MA321IGBa
PLACEMENT OF THE TRANSDUCERS
On the transducers is signed a side with the written N (North) and on the opposite side S
(South).
Put the transducer / transducers in correspondence of the part or parts to treat, if it is necessary
use the supplied elastic bands.
The different magnetic fields produced changing the transducers’ positions are shown in the
following figures. We have represented a leg, but obviously the connections are similar for
other parts of the body.
1 – Use of the two transducers to produce a coherent concatenate field. A transducer in contact
with the N part and the second with the S. The part to treat is between the two magnets.
Produced field:
COHERENT.
2 – Use of the two transducers to create a repelling field. The two transducers have the same
polarity in contact with the part to treat. The part to treat is between the two magnets. Produced
field:
REPELLING
3 – Use of the two transducers for therapy of two parts or greter parts. the two transducers are
put with the same part in contact. The part to treat is between the two magnets. Produced field:
NORMAL-WIDE.
4 – Use of one transducer. The transducer is directly put on the part to treat. Produced field:
NORMAL.