background image

ECG Palmare Cardio A

Cardio-A Palm ECG

ECG de poche Cardio A

ECG Palmar Cardio A

M33260-M-Rev

.3.02.17

MANUALE D’USO E MANUTENZIONE

USE AND MAINTENANCE BOOK

INSTRUCTIONS DE FONCIONNEMENT ET ENTRETIEN

MANUAL DE USO Y MANTENIMIENTO

0476

Fabbricante

Manufacturer

GIMA Spa

Via Marconi, 1 - 20060

Gessate (MI) - Italia - Made in P.R.C.

ATTENZIONE:

 Gli operatori devono leggere e capire completamente questo manuale 

prima di utilizzare il prodotto.

ATTENTION:

 The operators must carefully read and completely understand the present 

manual before using the product.

AVIS:

 Les opérateurs doivent lire et bien comprendre ce manuel avant d’utiliser le produit.

ATENCIÓN:

 Los operadores tienen que leer y entender completamente este manual 

antes de utilizar el producto.

Gima S.p.A. - Via Marconi, 1 - 20060 Gessate (MI) Italy

[email protected] - [email protected]

www.gimaitaly.com

Содержание Cardio-A Palm ECG

Страница 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Страница 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Страница 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Страница 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Страница 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Страница 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Страница 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Страница 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Страница 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Страница 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Страница 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Страница 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Страница 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Страница 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Страница 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...