4.3 Power

Power: 2×AAA alkaline batteries

Supply voltage range: 2.7~3.3VDC

Max average working current: ≤50mA (backlight off)

Auto power-off: 35s±5s (without operation)

Low battery indication: 2.5VDC±0.2VDC

4.4 Data Memory

- Data memory of up to about 30 seconds of ECG for every piece of record.

- Ability to Store 600 pieces of record at most.

- Able to transfer data records to the computer by data interface port. (Optional)

4.5 Overall Dimension and Weight

Dimensions: (L)115×(W)58×(H)18 mm

Weight: 100g (including 2 AAA batteries)

4.6 Classification

- The type of protection against electric shock: Internally powered equipment

- The degree of protection against electric shock: Type BF applied part

- The degree of protection against harmful ingress of water: Ordinary equipment without protection against ingress of water.

- Electro-Magnetic Compatibility: Group I, Class B

5 Accessories

A hanging cord

Two batteries (AAA)

A pouch

A User Manual

A Data Cable (optional)

An ECG lead wire (optional)

Note: The accessories are subject to change. Detailed items and quantity see the Packing List.

6 Maintenance and Service

6.1 Maintenance 

Please change the batteries, when appears low-voltage icon.

It is recommended to wipe the metal electrode with medical alcohol, when the measuring signal is too weak.

Please wipe the surface of the unit with clean cloth or let it dry in air.

Please take out the batteries if the unit will not be used for a long time.

Storage and Transportation

Ambient temperature:   

-20 to 60ºC

Relative humidity: 

 

10~ 95%

Atmospheric pressure:  

50~107.4KPa

This device should be transported by land (vehicle or railway) or air in accordance with the contractual 

terms. Do not hit or drop it with force and protect it from sunlight and rain.

Keep it away from corrosive substances, explosive substances, high/low temperature and moisture.

If the device gets wet or has water steam, please stop operating it.

When it is taken from cold environment to warm and humid environment, please DO NOT use it immediately.

DO NOT operate the button on front panel with sharp materials.

DO NOT immerge the device into liquid. Do NOT spray any liquid on it directly.

The life of this unit is 5 years. In order to ensure its long service life, please pay attention to the use of 

maintenance.

6.2 Cleaning and Disinfecting Instruction

Surface-clean sensor with a soft gauze by saturating with a solution such as 75% isopropyl alcohol, if low-level disinfection 

is required, use a 1:10 bleach solution. Then surface-clean with a damp cloth and dry with a piece of cloth.

Caution: Do not sterilize by irradiation steam, or ethylene oxide.

ENGLISH

Содержание Cardio-A Palm ECG

Страница 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Страница 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Страница 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Страница 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Страница 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Страница 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Страница 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Страница 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Страница 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Страница 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Страница 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Страница 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Страница 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Страница 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Страница 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...

Отзывы: