Accessories
Standard Configure:
Adult Cuff: limb circumference 22-32
cm (upper arm center)
User Manual, four “AA” alkaline batteries
AC Adapter:
Input: voltage: AC 100 V~240 V
frequency: 50
Hz/60 H Rated current: AC 150 mA
Output: DC 5.0
V±0.2
V
1.0
A
APPENDIX
Guidance and manufacturer’s declaration – electromagnetic emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer of the user of the device should
assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emission
CISPR 11
Class B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±
2 kV for power supply lines
±
2kV for power supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV differential mode
±
2 kV common mode
±
1 kV differential mode
±
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)for
0.5 cycle
40% U
T
(60% dip in U
T
)for 5
cycles
70% U
T
(30% dip in U
T
)for 25
cycles
<5% U
T
(>95% dip in U
T
)for
5 sec
<5% U
T
(>95% dip in U
T
)for
0.5 cycle
40% U
T
(60% dip in U
T
)for
5 cycles
70% U
T
(30% dip in U
T
)for
25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. The device
can continue the operation during
power mains interruptions due to
the usage of battery.
Power frequency (50/60Hz)
magnetic field
IEC61000-4-8
3A/m
3A/m
Mains power quality should be
that of a typical commercial or
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity –for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING