21
ENGLISH
Err09
The pulse signal is too weak or the cuff is loose.
Wrap the cuff correctly (refer to Chapter 10)
Err10
Cuff is blocked or squeezed
Wrap the cuff correctly (refer to Chapter 10)
Err14
Cuff leakage
Replace with a new cuff
Err11
Err12
Err13
The signal amplitude is too big owing to the arm or
body moving or other reasons when measuring
Keep arm, body still, measure again
Err16
Err19
It takes too much time
CHAPTER 11 - TROUBLESHOOTING
Abnormal
Phenomenons
Causes
Solutions
BP measurement
values too high or
too low.
Cuff is not connected correctly.
Correctly connect cuff.
Talk or move arm in measurement
Keep quiet and restart a measurement.
The turnup close oppress the arm
Take off the clothes, and restart a measurement
No pressure
Cuff leakage
Buy a new cuff.
The cuff windpipe is not correctly connected
with cuff
Correctly connect.
Cuff not inflate
Contact us.
Cuff deflate in short
time
Loose cuff
Correctly tangle cuff.
It can not carry on measurement ,even if press the measurement button Return on the power and restart a measurement.
Abruptly turn the
power off in adding
pressure
No use for a long time, the batteries can be
exhausted owing to the changed temperature
Replace all four batteries with new ones.
Hold the on/off
button but can not
start the device
Batteries are worn
Replace all four batteries with new ones.
The battery polarities is reversed
Check the battery installation for proper placement of the battery
polarities.
Cuff inflation start before press the measurement button
Stop using the device and contact us.
Cuff never deflation
Stop using the device and contact us.
Air pressure error
Deflation error
Pull out the cuff to deflate. Stop using the device and contact us.
Others
Keep arm, body still, measure again.
No press value displayed or the value unaltered when cuff inflating
Pull out the cuff to deflate. Stop using the device and contact us.
Other phenomenon
Switch on the power once again and restart an operation.
Replace the batteries.
If no, please contact us.
CHAPTER 12 - MAINTENANCE, CLEANING AND KEEPING
*Please do obey the precautions and correct operating methods in this user manual. Otherwise, we will not responsible for any fault.
Warning
Remove the batteries before cleaning. The accessories and main unit must be separated for cleaning.
Maintenance is not allowed during device using.
Do not squeeze the rubber tube on the cuff.
Caution
High pressure disinfection to the device and accessories is not allowed.
Do not let water or cleaning agent flow into the socket to avoid device damage.
Do not soak the device and accessories in liquid.
If any damage or deterioration of the device and accessories is found, please do not use it.
Maintenance:
Clean the device and accessories regularly. It is recommended to clean them every one month. When the device or accessory gets dirty, use a dry
and soft cloth to wipe. If they are very dirty, it is available to dip the soft cloth into water or mild detergent, and wring out, then use the cloth for
cleaning.
The device shall be inspected and calibrated regularly (or according to inspection standard of hospital). The inspection can be carried out in