ENGLISH
Your device may not contain all the following symbols.
Signal
Description
Signal
Description
Caution: read instructions (warnings) carefully
Follow instructions for use
SYS
Systolic pressure
DIA
Diastolic pressure
MAP
Mean blood pressure
PUL (Pressione
del polso)
Pulse rate (bpm)
IPXX
Enclosure protection grade
EMC
Electromagnetic compatibility
Recyclable
P/N
Material code of manufacturer
Lot number
Use by date
This way up
Fragile, handle with care
Keep in a cool, dry place
Atmospheric pressure limit
Temperature limit
%
Humidity limit
Manufacturer
Date of manufacture
Batteries Power
Serial number
F
lating
D
eflating
WEEE disposal
Type BF applied part
Serial number
Medical Device complies with Directive 93/42/EEC
Authorized representative in the European community
Class II applied
Socket for power adapter
Interface for connecting cuff
Voice deactivation
Voice activation
Great movement when measuring
Cuff tied correctly
Artery indicator label
CHAPTER 10 - ERROR MESSAGE
When the high pressure position appears “Err” and the low pressure position appears the error number, the measurement is not normal.(Error
number are 02,04,06-16,19)
Error Mark
Causes
Solutions
Err02
Err15
Function abnormal
Please contact us
Err04
Low battery
Please replace the battery or link adapter
Err06
The cuff is not wrapped correctly.
Wrap the cuff correctly (refer to Chapter 10)
Err07
Cuff plug fall off
Make sure the cuff plug is securely inserted in the windpipe (refer to
Chapter 10)
Err08
Air pressure error
Keep arm, body still, measure again