22
ENGLISH
appointed institutions, or by professional personnel or contact us for inspection. Under the setting interface, Press the “START/STOP” button once,
after “CAL” appears on the screen, press “M” button for more than 15 s to enter the static pressure interface.
Advice
Do not use gasoline, volatile oil, diluent, etc. to wipe the device.
Do not clean or wet the cuff.
Storage:
Advice
Do not expose the device in direct sunlight for long time, otherwise the display screen maybe damaged.
The basic performance and safety of the device are not affected by the dust or cotton wool in home
environment,while the device shall not be placed where with high temperature, humidity or dusty.
Aged cuff may result in inaccurate measurement, please replace the cuff periodically according to the user manual.
To avoid device damage, keep the device out the reach of children and pets.
Avoid the device close to extreme high temperature such as fireplace, otherwise the device performance may be affected.
Do not store the device with chemical medicine or corrosive gas.
Do not place the device where there is water.
Do not place the device where with slope, vibration or impact
Take the batteries out if the device is not to be used for three months or longer.
CHAPTER 13 - NIBP SPECIFICATION
Name
Electronic Sphygmomanometer
The degree of protection against ingress of water
IPX1
Display
LED
NIBP Specifications
Measurement Method
Oscillometric method
Working mode
Automatic
Operation mode
Continuous operation
Pressure Range
0~297 mmHg(0~39.6 kPa)
Measurement range
Pressure
SYS
:
30
~
270 mmHg(4
~
36 kPa)
DIA
:
10
~
220 mmHg(1.3
~
29.3 kPa)
Pulse: 40~240/min
Inflation
160±5 mmHg(21.33±0.67 kPa)
Overpressure protect
297±3 mmHg(39.6±0.4 kPa)
Resolution
Pressure: 1
mmHg(0.133 kPa)
Accuracy
Static pressure: ±3 mmHg(±0.4 kPa)
Error
The BP value measured by the device is equivalent with the measurement value of
Stethoscopy, perform clinical verification in accordance with the requirements in ISO
81060-2: 2013, whose error meets the followings:
Maximum mean error: ±5 mmHg
Maximum Standard deviation: 8 mmHg
Operating Temperature/ Humidity
+5ºC~40 ºC . 15%RH~85%RH(no condensat
ion)
Transport
Transport by general vehicle or according to the order contract, avoid pounded, shake and
splash by rain and snow in transportation.
Storage
Temperature: -20 ºC~+55 ºC; Relative humidity: ≤95
%; No corrosive gas
and
drafty
.
Atmospheric pressure
700 hPa~1060 hPa
Power supply
4 “AA” alkaline batteries, AC Adapter(AC, 100 V-240 V, opti
onal)
Rated current
≤ 600 mA
Battery life
When the temperature is 23 ºC, limb circumference is 270 mm, the measured blood
pressure is normal, 4 “AA” alkaline batteries cab be used about 300 times.
Main Unit Dimensions
129
*
101
*
72 mm
Main Unit Weight
281 gram(without batteries)
Safety classification
Class
Ⅱ
equipment (power supplied by power adapter)/Internally powered equipment
(power supplied by batteries)
Type BF applied part
Service life
The service life of the device is five years or 10000 times of BP measurement.
Date of manufacturer
See the label