24
•
: key-lock icon, when this icon appears, it means that the key operation is disabled.
Note:
the key-lock status can be set at any screen view by combining the given keys. During key-lock status,
all key operations are disabled except Power button and combination keys for unlocking operation.
•
:
net connection icon, indicates that the device is connected to the network. If the device is discon-
nected from the network, then the icon will disappear
•
:
battery voltage indicator.
•
PLETH
: indicating the displayed waveform is plethysmogram.
Waveform area:
•
II:
ECG lead.
•
X1:
ECG waveform gain.
•
PI% 12.3:
the label and value of perfusion index. This item is displayed only when “Setup Menu
SpO
2
PI Display” is set as “ON.
Note:
PI function is optional, please refer to the Gima Vital Signs Monitor in your hand, we will not cover it
again.
•
37.5°C/-- °C:
the measured temperature value. This function is optional.
If the accessories are connected incorrectly or disconnected from the Gima Vital Signs Monitor, the message
“Probe off” and (or) “ECG lead off” will appear on the screen
NIBP list area:
When blood pressure measurement is taken, the data display area displays the recent 4 groups of data, the
form is as shown in figure 4.2.
•
Time:
the measuring time.
•
SYS/DIA/MAP:
systolic/diastolic/arterial mean pressure.
•
PR:
the measured pulse rate from the blood pressure measuring channel or the pulse rate value from oxime-
try measuring channel. The PR value from oximetry will take priority in the display
•
SpO
2
:
oxygen saturation (SpO
2
for short).
•
TEMP:
the temperature value.
•
HR:
the heart rate.
Note:
1. If the device is re-started, the data in the list data area will be cleared.
2. Invalid values will be displayed as “--"
Prompt Info. area:
•
MAP over-limit:
displays a message for current alarm event indicating that the measured MAP value exceeds
the preset value.
•
Mute 112
:
display the status of the alarm silence, and the counting down the time that the alarm sound
is silenced for.
shows that the alarm sound is enabled;
shows that the alarm sound is silent temporarily for 120
seconds;
shows that the alarm sound is disabled.
Operation instruction:
• Short pressing the display view
key to shift screen views.
• Long press the display view
key to enter into Menu setup screen.
• For Gima Vital Signs Monitor without ECG function, hold Auxiliary key
, firstly, then press NIBP setup
key
by; doing this it can lock / unlock key operation. For monitor with ECG function, hold ECG lead
key
, firstly, then press NIBP setup key
; by doing this it can lock / unlock key operation.
Nota:
This function is available in any screen view, we will not cover it again in the following