11.2 Factory Default Alarming Values and Setup Range
The factory default alarming value:
Intervalli di impostazione valore alto e basso:
Note:
it's limited for the high/low alarm setting value, which ensure that the high-limit setting value will not less
than (or equal to) that of the low-limit setting value. See the above table for detailed setting step.
Mode
Parameter
Adult
Pediatric
Neonate
HR
High limit
180bpm
200bpm
220bpm
Low limit
40bpm
50bpm
50bpm
SYS
High limit
180mmHg
130mmHg
110mmHg
Low limit
60mmHg
50mmHg
50mmHg
DIA
High limit
120mmHg
90mmHg
90mmHg
Low limit
50mmHg
40mmHg
30mmHg
MAP
High limit
160mmHg
110mmHg
100mmHg
Low limit
50mmHg
40mmHg
30mmHg
SpO
2
High limit
100%
100%
100%
Low limit
90%
85%
85%
Pulse rate
High limit
180bpm
200bpm
220bpm
Low limit
40bpm
50bpm
50bpm
TEMP
High limit
39,0°C
39,0°C
39,0°C
Low limit
35,0°C
35,0°C
35,0°C
Mode
Parameter
Adult
Pediatric
Neonate
Fase di
impostazione
HR
High limit
(1~350) bpm
(1~350) bpm
(1~350) bpm
1bpm
Low limit
(0~349) bpm
(0~349) bpm
(0~349) bpm
1bpm
SYS
High limit
(30~280) mmHg
(30~200) mmHg
(30~135) mmHg
1mmHg
Low limit
(29~279) mmHg
(29~199) mmHg
(29~134) mmHg
1mmHg
DIA
High limit
(11~232) mmHg
(11~150) mmHg
(11~100) mmHg
1mmHg
Low limit
(10~231) mmHg
(10~149) mmHg
(10~99) mmHg
1mmHg
MAP
High limit
(21~242) mmHg
(21~165) mmHg
(21~110) mmHg
1mmHg
Low limit
(20~241) mmHg
(20~164) mmHg
(20~109) mmHg
1mmHg
SpO
2
High limit
1~100%
1~100%
1~100%
1%
Low limit
0~99%
0~99%
0~99%
1%
Pulse rate
High limit
(1~300) bpm
(1~350) bpm
(1~350) bpm
1bpm
Low limit
(0~299) bpm
(0~349) bpm
(0~349) bpm
1bpm
TEMP
High limit
(0,1~60)°C
(0,1~60)°C
(0,1~60)°C
0,1°C
Low limit
(0~59,9)°C
(0~59,9)°C
(0~59,9)°C
0,1°C