User Manual for Vital Signs Monitor
28
“Store”:
decide whether to store ECG data or not. If your selection is “off”, thereafter the ECG data
measured will not be stored.
“T probe”:
the type of temperature probe “KRK”.
4.7.2 SpO
2
Setup
Figure 4.12 SpO
2
Setup Screen
Screen Description:
“SpO
2
”:
SpO
2
alarm switch; “
” indicates SpO
2
alarm is on; “
” indicates SpO
2
alarm is off.
“SpO
2
Hi”:
high limit of SpO
2
alarm; range: “1~100”.
“SpO
2
Lo”:
low limit of SpO
2
alarm; range: “0~99”.
“PR
”:
pulse rate alarm switch; “
” indicates PR alarm is on; “
” indicates PR alarm is off.
“PR Hi”:
high limit of PR alarm; range: “22~250”.
“PR Lo”:
low limit of SpO
2
alarm; range: “0~248”.
“PI display”:
“on” means PI display is enabled; “off” means PI display is disabled.
Operation Instructions
1. Press “
▲
” key or “
▼
” key to move cursor to select parameter.
2. Press “
■
” key to confirm and active this parameter setting.
3. Press “
▲
” key or “
▼
” again to adjust or modify parameter value.
4. Press “
■
”key again to confirm and save the setting.
5. Press “
” key to return to upper level screen.