User Manual for Vital Signs Monitor
13
Adult
300mmHg
Child
240mmHg
Neonate
140mmHg
Table B
After the verification, do press the button again to return to normal working mode, then
continue other operation, or the NIBP key will be invalid.
Pressure accuracy verification must be operated by technician or equipment manager. Doctor
or nurse is not allowed to do the verification, it is very dangerous especially when the pressure
cuff is still on patients.
¾
Air Leakage Check
In order to avoid significant error of blood pressure measurement or even no measurement result caused
by air leakage in the pneumatic system including the cuff during measuring, it is recommended to check if
there is leak in the pneumatic system as well.
Please remove the cuff from patient while performing the leakage check.
Safety Instructions for NIBP Monitoring
0
When taking the measurement of a pediatric or an infant or neonate’s (less than 10 years old)
blood pressure, do NOT operate in the adult mode. The high inflation pressure may cause
lesion or even body putrescence.
0
It is recommended to take the blood pressure measurement manually. Automatic
measurement should be used at the presence of a doctor/nurse.
0
NIBP monitoring is prohibited to those who have severe hemorrhagic tendency or with sickle
cell disease, or partial bleeding will appear.
0
Pay attention to the color and sensitivity of the limb when measuring NIBP; make sure the
blood circulation is not blocked. If blocked, the limb will discolor, please stop measuring or
remove the cuff to other positions. Doctor should examine this timely.
0
Confirm your patient category (adult, pediatric or neonate) before measurement.
0
Do NOT bind NIBP cuff on limbs with transfusion tube or intubations or skin lesion area,
otherwise, damages may be caused to the limbs.
0
If the time of the automatic pattern noninvasive blood pressure measurement is too long, the
body connected with the cuff will possibly occur the purpura, lack the blood and the
neuralgia. In order to protect patient, it is requested to inspect the luster, the warmth and the
sensitivity of the body far-end frequently. Once observes any abnormity, please immediately
stop the blood pressure measurement.
0
The patient should lie on the back so that the cuff and the heart are in a horizontal position
and the most accurate measure is taken. Other postures may lead to inaccurate measurement.
0
Do not speak or move before or during the measurement. Ensure that the cuff will not be hit
or touched by other objects.