User Manual for Vital Signs Monitor
5
2 SYS:
display systolic pressure value.
3 DIA:
display diastolic pressure value.
4 MAP:
When NIBP measurement mode is set to “manual” and “STAT”: Display mean arterial
pressure or measuring time of the latest group of NIBP measurement; they will be displayed
alternately. The format of NIBP measuring time is “hh:mm”. If the tourniquet is in use, the cuff
pressure will be displayed here; When NIBP measurement mode is set to “AUTO”: Display
real-time pressure value during measurement. Countdown time will be displayed in the MAP when
the measurement finishes. Countdown time has two formats (
>
1 hour HH: mm;
<
1 hour mm: ss).
Note:
two formats to display NIBP value: “×××mmHg” and “××.×kPa”. Refer to section “4.4.2 NIBP
Setup” to set the unit of NIBP value; the conversion relation between “mmHg” and “kPa”:
1mmHg=0.133kPa.
5 HR(priority indicator)
: if HR indicator is on, it indicates that the numerical value beside is HR
measuring value;
6 Display HR or PR value:
when the set of “Setup Menu
→
System
→
priority” is “HR”, it shows HR
value here preferentially; if the set is “PR”, PR value will be shown preferentially.
7 PR(priority indicator):
if PR indicator is on, it indicates that the numerical value beside is pulse
rate value; Unit: “bpm (beats per minute)”.
If SpO
2
and NIBP are monitoring at the same time, only the PR value measured through SpO
2
probe
is displayed here and the PR measuring by NIBP measurement is undergoing and recorded in NIBP
list.
8 SpO
2
:
Display SpO
2
value; Unit: “%”.
9 “ ”:
Bar-graph of pulse intensity.
10 LCD panel
11 Pulse sync indicator patient category indicator:
“
” for adult; “ ” for pediatric; “
” for
neonate; Patient category is selected under sub-menu “Patient Info” within the setup menu.
12 Pulse sync indicator
: Cardio-pulse/pulse sync indicator. When HR priority indicator is on, its
flashing is synchronized with heart beat; When PR priority indicator is on, its flashing is
synchronized with pulse.
13
NIBP:
start/cancel NIBP measurement.
14
Alarm silence key:
Enable/disable alarm silence function. When the alarm silence indicator
on the left of keys is on, it means the system is in alarm silence status and it lasts this status for 2
minutes. When finishing counting down, the system will resume normal alarm status automatically,
if alarm event occurs at this time the alarm sound will be effective again.
0
DO NOT silence the audible alarm or decrease its volume if patient safety could be
compromised.