18
Figure 6-4 Installation of limb electrodes
6.4.3 Colors of lead cables
As shown in Table 6-1:
Table 6-1 Colors of lead cables
Electrode position
European standard
American standard
Mark
Color
Mark
Color
Right arm
R
Red
RA
White
Left arm
L
Yellow
LA
Black
Left leg
F
Green
LL
Red
Right leg
N/RF
Black
RL
Green
Chest 1
Cl
Red
Vl
Red
Chest 2
C2
Yellow
V2
Yellow
Chest 3
C3
Green
V3
Green
Chest 4
C4
Brown
V4
Blue
Chest 5
C5
Black
V5
Orange
Chest 6
C6
Purple
V6
Purple
Note
n
It is recommended to install the lead cables after turning off the device.
n
Apply appropriate amount of conductive paste on the electrode when installing the
electrode.
n
If the ECG waveform does not appear for a long time, check if the electrode is in
good contact with the skin.