57
Note:
The heart abnormalities such as posterior myocardial ischemia, early posterior MI and old
posterior MI are not included in the database. These abnormalities and other heart disorders not
contained in above sheet won’t be regarded as the judgment object for the verification of
automated interpretation accuracy.
4.6 Data preprocessing
4.6.1 CTS preprocessing
The 16 cases (CAL05000, CAL10000, CAL15000, CAL20000, CAL20002, CAL20100,
CAL20110, CAL20160, CAL20200, CAL20210, CAL20260, CAL20500, CAL30000, ANE20000,
ANE20001, ANE20002) from CTS-ECG shall be processed for voltage conversion and frequency
conversion for resampling as the applicable format in the system. Then cases will be imported to the
device. After that, the verification of automated measurement parameters will be carried on.
4.6.2 CSE preprocessing
The cases (MA_0001~MA0125, D_0001~D_1220) from the CSE shall be processed for
voltage conversion and frequency conversion for resampling as the applicable format in the
system. Then cases will be imported to the device. After that, the case of MA_0001~MA0125
shall be used for the following verification of automated measurement parameters, and the case
Содержание 1200G ECG