30
7.7 Sampling setup
In the main interface, click
to enter the sampling setup interface. The optional
content of each setting item and its description are shown in the following table:
Item
Options
Description
AC Filter
ON/OFF
Turn on or off the AC filter.
EMG Filter
ON/OFF
Turn on or off the EMG filter.
DFT Filter
ON/OFF
Turn on or off the Baseline filter.
Show Style
[1leads]/[3leads]/[6leads]/
[12 leads]
Set the display method of ECG.
Sort Lead
[Routine Lead]/[Cabrera
Lead]
Set the arrangement of the leads.
Show Gain
[2.5mm/mV]/[5mm/mV]/[10
mm/mV]/[20mm/mV]/[40m
m/mV]
Set the gain of displayed ECG.
Show Speed
[5mm/s]/[6.25mm/s]/[10mm
/s]/[12.5mm/s]/[25mm/s]/[50
mm/s]
Set the speed of displayed ECG.
Background
Grid
[Show]/[Not Show]
Set to use the background grid or not.
Premature
0~100
The system will use the input value as a
standard of judging premature beat.
Pause Time
1200~3000
The system will use the input value as a
standard of judging beat pause.
Tachycardia
0
~
250
The system will use the input value as a
standard of judging tachycardia.
Bradycardia
0
~
99
The system will use the input value as a
standard of judging bradycardia.
Default
All above settings will restore to default after clicking this button.
7.8 Print setup
In the main interface, click
to enter the print setup interface. The optional
content of each setting item and its description are shown in the following table:
Item
Options
Description
Print Mode
[Auto 12
×
1]/[Auto
6
×
2+1]/[Auto
6
×
2+1_H]/[Auto 6
×
2]/[Auto
6
×
2_H]/[Auto 3
×
4+1]/[Auto
3
×
4+2]/[Auto
The system takes the selected option as
default print mode.