23
Display field
The screen displays sampled 12-lead ECG waveform, by double clicking the waveform,
you can switch between 1-lead, 3-lead, 6-lead and 12-lead.
Operation field
Control the print display mode of the device through corresponding operation settings.
1.
Speed: use the
button to switch the speed between 5 mm/s, 6.25 mm/s, 10 mm/s,
12.5 mm/s, 25 mm/s and 50 mm/s.
2.
Gain (sensitivity): use the
button to switch the gain between 2.5 mm/mV, 5
mm/mV, 10 mm/mV, 20 mm/mV and 40 mm/mV. The overall gain (sensitivity) can be checked
by calibration function.
3.
Filter: use the
button to switch the filter between NONE, AC, EMG, AC+EMG,
DFT, AC+DFT, EMG+DFT and AC+EMG+DFT.
In which: AC AC filter
EMG
EMG filter
DFT
Baseline filter
4.
Print mode: in print setup, when the data type is set to “After Print”, use the
button
to switch the print mode between Manual, Auto 12
×
1, Auto 6
×
2+1, Auto 6
×
2+1_H, Auto 6
×
2,
Auto 6
×
2_H, Auto 3
×
4+1, Auto 3
×
4+2, Auto 1
×
12, Rhythm 12, Rhythm 10, Rhythm 8 and
Rhythm 6.
5. Print/End print: use the
button to start or end the printing operation.
1)Auto mode: After starting to print, the system automatically prints and stores the real-time
12-lead ECG waveform. The length is determined by the relevant settings in the print setup.
Based on the settings, the automatic analysis data and conclusions are printed, and the system
automatically ends printing.
2)Manual mode: After starting to print, user need to switch the lead to print the waveform of
different leads, that is, the ECG printed in the manual mode is asynchronous, and the data is not
saved. User need to press the PRINT button again when the print needs to be terminated.
3)If lead-off occurs during sampling process, the printed waveform will be marked with “*”.
4)If lead overload occurs during sampling process, the printed waveform will be marked with
“+”.
6. During sampling process, press the PRINT button and wait until the printing is competed, it
will pop up a dialog box, which includes “Delete check box”, “Review”, “OK” and prompts (If
lead-off or overload occurs during sampling process, there will be a prompt in this dialog box).
Clicking “Review” could enter the case management to view sampled cases; Select the delete
check box and click “OK” to delete the case.
7. End sampling: After the device starts sampling, use the
button to end the sampling,