12 MAINTENANCE
You may perform the following:
- clean or autoclave the parts specifi ed under
“Cleaning and Decontamination”,
- replace the parts specifi ed under “Spare Parts”,
- lubricate the pistons,
- change the battery,
- calibrate and readjust the pipette.
쎱
Operation for Single-channel models
Tip-holder and Tip-ejector
These parts must be changed, if they are accidentally
damaged or attacked chemically. You should also re-
move these parts for cleaning or decontamination
purposes.
Changing the Tip-ejector
1) K e e p t h e t i p -
ejec tor button
depressed and
grip the top of the
tip-ejector with
the other hand.
Symptom
Possible Cause
Refer to page
“INIT1” on display
The pipette has not been locked after
6
10 minutes of inactivity.
Volume has not been set to the correct
29-31
calibration volume when the calibration
tool was plugged in.
“INIT 2” on display
Setting problems.
29-31
“INIT 3” on display
Battery was changed.
24-25
Other Error Message
If the following occur, contact your Gilson distributor.
Message Cause
ERR4
Electronic problems (microchip).
Tip-ejector