32
PIPETMAN
®
u
ltra
Each pipette is inspected and validated by qualifi ed
technicians according to the Gilson Quality System.
Gilson declares that its manufactured pipettes comply
with the requirements of the ISO 8655 Standard, by
type testing.
The adjustement is carried out under strictly defi ned
and monitored conditions (ISO 8655-6):
• Basis of adjustment, Ex.
• Reference temperature, 20 °C
• Relative humidity, 50 %
• Barometric pressure, 101 kPa
• Use of distilled water grade 3 (ISO 3696)
• Ten measurements for each test volume, which are
Nominal Volume, 50 % of Nominal Volume, and the
minimum or 10 % of Nominal Volume.
Maximum Permissible Errors
Model Volume
Gilson
ISO
8655
(Reference)
(μL)
Systematic
Random
Systematic
Random
Diamond Tip
error (μL)
error (μL)
error (μL)
error (μL)
8x20
(F21040)
Min.
1
± 0.10
압
0.08
± 0.40
압
0.20
and
2
± 0.10
압
0.08
± 0.40
압
0.20
12x20
(F21041)
10
± 0.20
압
0.10
± 0.40
압
0.20
DL10
Max
20
±
0.40
압
0.20
± 0.40
압
0.20
8x300
(F21042)
Min.
20
± 1.00
압
0.35
± 8.00
압
3.00
and
30
±
1.00
압
0.35
± 8.00
압
3.00
12x300
(F21043)
150
± 1.50
압
0.60
± 8.00
압
3.00
D300
Max.
300
±
3.00
압
1.00
± 8.00
압
3.00
The data given in the tables conform to the ISO 8655-2
Standard.