Maximum Permissible Errors

The data provided in the following tables conform to 

the IsO 8655-2 standard. 

*P5000 and P10ml do not have tip ejectors.

NOTICE

Do not lubricate this pipette. The use of 

lubricant cancels the warranty of this pipette.

VOlUME

(µl)

GILSON

ISO 8655

SYSTEMATIC 

ERROR (µL)

RANDOM 

ERROR (µL)

SYSTEMATIC 

ERROR (µL)

RANDOM 

ERROR (µL)

P2 (P/N F144801)

 

WITH D10 OR DL10 TIPS

0.2

± 0.024

 0.012

± 0.08

 0.04

0.5

± 0.025

 0.012

± 0.08

 0.04

1

± 0.025

 0.012

± 0.08

 0.04

2

± 0.030

 0.014

± 0.08

 0.04

P10 (P/N F144802) 

WITH D10 OR DL10 TIPS

1

± 0.025

 0.012

± 0.12

 0.08

5

± 0.075

 0.030

± 0.12

 0.08

10

± 0.100

 0.040

± 0.12

 0.08

P20 (P/N F123600) 

WITH D200 TIPS

2

± 0.10

 0.030

± 0.2

 0.1

5

± 0.10

 0.040

± 0.2

 0.1

10

± 0.10

 0.050

± 0.2

 0.1

20

± 0.20

 0.060

± 0.2

 0.1

P100 (P/N F123615)

 

WITH D200 TIPS

10

± 0.35

 0.10

± 0.8

 0.3

50

± 0.40

 0.12

± 0.8

 0.3

100

± 0.80

 0.15

± 0.8

 0.3

P200 (P/N F123601) 

WITH D200 TIPS

20

± 0.50

 0.20

± 1.6

 0.6

100

± 0.80

 0.25

± 1.6

 0.6

200

± 1.60

 0.30

± 1.6

 0.6

P1000 (P/N F123602)

 

WITH D1000 TIPS

100

± 3.0

 0.6

± 8.0

 3.0

500

± 4.0

 1.0

± 8.0

 3.0

1000

± 8.0

 1.5

± 8.0

 3.0

P5000* (P/N F123603)

 

WITH D5000 TIPS

500

± 12

 3

± 40

 15.0

2500

± 15

 5

± 40

 15.0

5000

± 30

 8

± 40

 15.0

P10mL* (P/N F161201) 

WITH D10ML TIPS

1000

± 30

 6

± 60

 30.0

2000

± 30

 6

± 60

 30.0

5000

± 40

 10

± 60

 30.0

10000

± 60

 16

± 60

 30.0

Содержание F123600

Страница 1: ...Quick Guide EN CLASSIC...

Страница 2: ...ting the Tips on page 5 Safety bag Certi cate of Conformity including barcode sticker GLP Compliant The serial number is engraved on the body of the pipette It provides unique identification of your p...

Страница 3: ...genuine PIPETMAN DIAMOND Tips are used Each pipette is inspected and validated by quali ed technicians in accordance with the Gilson Quality System Gilson declares that its manufactured pipettes comp...

Страница 4: ...N F144802 WITH D10 OR DL10 TIPS 1 0 025 0 012 0 12 0 08 5 0 075 0 030 0 12 0 08 10 0 100 0 040 0 12 0 08 P20 P N F123600 WITH D200 TIPS 2 0 10 0 030 0 2 0 1 5 0 10 0 040 0 2 0 1 10 0 10 0 050 0 2 0 1...

Страница 5: ...2 L 0 01 L 0 002 L P10 to P20 L 0 1 L 0 02 L P100 P200 L 0 2 L P1000 P5000 0 01 mL mL 0 002 mL P10mL mL 0 1 mL 0 02 mL Fitting the Tips To fit a new PIPETMAN DIAMOND Tip push the tip holder into the t...

Страница 6: ...smoothly to the top position to aspirate the set volume of liquid 4 Wait one second The time depends on the model refer to the table Then withdraw the pipette tip from the liquid You may wipe any dro...

Страница 7: ...bnormal use and or improper or inadequate maintenance contrary to the recommendations given in the user s guide including but not limited to pipettes which have been subjected to physical damage impro...

Страница 8: ...www gilson com contactus LT801623 C 2019 Gilson Inc All rights reserved...

Отзывы: