4
QUICK GUIDE
|
PIPETMAN® ClAssIC
Maximum Permissible Errors
The data provided in the following tables conform to
the IsO 8655-2 standard.
*P5000 and P10ml do not have tip ejectors.
NOTICE
Do not lubricate this pipette. The use of
lubricant cancels the warranty of this pipette.
VOlUME
(µl)
GILSON
ISO 8655
SYSTEMATIC
ERROR (µL)
RANDOM
ERROR (µL)
SYSTEMATIC
ERROR (µL)
RANDOM
ERROR (µL)
P2 (P/N F144801)
WITH D10 OR DL10 TIPS
0.2
± 0.024
≤
0.012
± 0.08
≤
0.04
0.5
± 0.025
≤
0.012
± 0.08
≤
0.04
1
± 0.025
≤
0.012
± 0.08
≤
0.04
2
± 0.030
≤
0.014
± 0.08
≤
0.04
P10 (P/N F144802)
WITH D10 OR DL10 TIPS
1
± 0.025
≤
0.012
± 0.12
≤
0.08
5
± 0.075
≤
0.030
± 0.12
≤
0.08
10
± 0.100
≤
0.040
± 0.12
≤
0.08
P20 (P/N F123600)
WITH D200 TIPS
2
± 0.10
≤
0.030
± 0.2
≤
0.1
5
± 0.10
≤
0.040
± 0.2
≤
0.1
10
± 0.10
≤
0.050
± 0.2
≤
0.1
20
± 0.20
≤
0.060
± 0.2
≤
0.1
P100 (P/N F123615)
WITH D200 TIPS
10
± 0.35
≤
0.10
± 0.8
≤
0.3
50
± 0.40
≤
0.12
± 0.8
≤
0.3
100
± 0.80
≤
0.15
± 0.8
≤
0.3
P200 (P/N F123601)
WITH D200 TIPS
20
± 0.50
≤
0.20
± 1.6
≤
0.6
100
± 0.80
≤
0.25
± 1.6
≤
0.6
200
± 1.60
≤
0.30
± 1.6
≤
0.6
P1000 (P/N F123602)
WITH D1000 TIPS
100
± 3.0
≤
0.6
± 8.0
≤
3.0
500
± 4.0
≤
1.0
± 8.0
≤
3.0
1000
± 8.0
≤
1.5
± 8.0
≤
3.0
P5000* (P/N F123603)
WITH D5000 TIPS
500
± 12
≤
3
± 40
≤
15.0
2500
± 15
≤
5
± 40
≤
15.0
5000
± 30
≤
8
± 40
≤
15.0
P10mL* (P/N F161201)
WITH D10ML TIPS
1000
± 30
≤
6
± 60
≤
30.0
2000
± 30
≤
6
± 60
≤
30.0
5000
± 40
≤
10
± 60
≤
30.0
10000
± 60
≤
16
± 60
≤
30.0