PIPETMAN® ClAssIC

Introduction

 

PIPETMAN® Classic is a fully adjustable, air-displacement 

pipette that is used with disposable tips.
Eight single channel models cover a volume range from 

0.2 µl to 10 ml for many applications.

Parts Checklist

 

PIPETMAN Classic

 

PIPETMAN Classic Quick Guide

 

Dual-position adapter (for P2 and P10 models only; 

refer t

Fitting the Tips  

on page 5)

 

safety bag

 

Certificate of Conformity (including barcode 

sticker)

GLP Compliant

The 

serial number

 is engraved on the body of the 

pipette. It provides unique identification of your pipette 

and the date of manufacture. For more details, refer to 

PIPETMAN Classic User’s Guide lT801120. Ex: 

QG10369

The 

barcode

 on the box and the Certificate of 

Conformity provide traceability of your pipette.

Figure 1

serial number location

NOTE

The updated line of PIPETMAN Classic 

is identified with an underlined serial 

number starting from 

QG

.

serial Number

Содержание F123600

Страница 1: ...Quick Guide EN CLASSIC...

Страница 2: ...ting the Tips on page 5 Safety bag Certi cate of Conformity including barcode sticker GLP Compliant The serial number is engraved on the body of the pipette It provides unique identification of your p...

Страница 3: ...genuine PIPETMAN DIAMOND Tips are used Each pipette is inspected and validated by quali ed technicians in accordance with the Gilson Quality System Gilson declares that its manufactured pipettes comp...

Страница 4: ...N F144802 WITH D10 OR DL10 TIPS 1 0 025 0 012 0 12 0 08 5 0 075 0 030 0 12 0 08 10 0 100 0 040 0 12 0 08 P20 P N F123600 WITH D200 TIPS 2 0 10 0 030 0 2 0 1 5 0 10 0 040 0 2 0 1 10 0 10 0 050 0 2 0 1...

Страница 5: ...2 L 0 01 L 0 002 L P10 to P20 L 0 1 L 0 02 L P100 P200 L 0 2 L P1000 P5000 0 01 mL mL 0 002 mL P10mL mL 0 1 mL 0 02 mL Fitting the Tips To fit a new PIPETMAN DIAMOND Tip push the tip holder into the t...

Страница 6: ...smoothly to the top position to aspirate the set volume of liquid 4 Wait one second The time depends on the model refer to the table Then withdraw the pipette tip from the liquid You may wipe any dro...

Страница 7: ...bnormal use and or improper or inadequate maintenance contrary to the recommendations given in the user s guide including but not limited to pipettes which have been subjected to physical damage impro...

Страница 8: ...www gilson com contactus LT801623 C 2019 Gilson Inc All rights reserved...