9
3 - RECOMMENDATIONS
The pipette can be used b 4 °C
and + 40 °C but the specifications may vary
according to the temperature (see Chapter
4 for controlled conditions of use).
When the
red flag appears, be prepared to refill
the DistriTip before you reach the last aliquot.
For the complete distribution of an aliquot, you must
press the push-button until it reaches the adjust-
ment ring. If after distributing an aliquot, a drop of
liquids forms on the end of the DistriTip, do not
discard this drop, which is part of the next aliquot.
Accuracy could be affected by capillary action;
when dispensing, do not ‘wipe’ the DistriTip against
the side of the recipient vessel. After changing the
DistriTip or resetting the volume, you may want to
discard the first aliquot.
Never try to force the volumeter beyond the recom-
mended volume ranges.
4 - SPECIFICATIONS
The data given in the table conform to the ISO
8655-2 Standard.
Maximum Permissible Errors
DistriTips
model
Volume
Gilson
ISO 8655
(Reference)
(µL) Systematic Random Systematic Random
error (µL) error (µL) error (µL) error (µL)
125 µL
Min.
2 ± 0.100
0.080 ± 0.20
0.10
Micro
(F164100)
5 ± 0.125
0.075 ± 0.20
0.10
Micro ST
(F164130)
Max 10 ± 0.200
0.100 ± 0.20
0.10
1250 µL
Min. 20 ± 0.80
0.20 ± 1.50
0.60
Mini
(F164110)
50 ± 1.00
0.40 ± 1.50
0.60
Mini ST
(F164140)
Max. 100 ± 1.00
0.60 ± 1.50
0.60
12.5 mL
Min. 200 ± 6.00
1.00 ± 12.00
4.00
Maxi
(F164120)
500 ± 7.50
1.50 ± 12.00
4.00
Maxi ST
(F164150)
Max. 1000 ± 10.00
2.50 ± 12.00
4.00
ST means Sterilized.