Lock the selected volume 

So that the volume cannot be changed accidentally, 
you must push the black adjustment ring down 
firmly
 to lock the selected volume. You hear a click 
when you lock the volume.

When you want to change the volume, you should 
unlock the adjustment ring by pulling it upwards.

If you forget to unlock it, the action of turning the 
adjustment ring will not harm the mechanism.

Filling the syringe 

You only need one hand to fill the 
DistriTip.  By  simply  pushing  the 
filling-knob  with  your  thumb,  you 
may  completely  or  partially  fill  the 
DistriTip, which is clearly marked with 
graduations according to its size.
1-  Push the filling-knob downwards 

to  its  lowest  possible  position, 
using your thumb.

2-  Immerse the end of the DistriTip 

into  the  liquid  to  be  aspirated. 
When aspirating, ensure that the 
tip remains below the surface of 
the liquid (about 2 mm).

3-  Move  the  filling-knob  slowly  upwards  to  its 

highest position to completely fill the syringe, 
or  with  reference  to  the  graduations  on  the 
DistriTip, to an intermediate position.

4-  Withdraw the DistriTip from the liquid and care-

fully wipe any excess liquid from the outside, 
using a medical wipe, 

taking care not to touch 

the orifice of the syringe

After fitting a new DistriTip, you are advised to 
‘wet’ the interior with the liquid to be aspirated. 
Then purge and refill the syringe. 

Dispensing the aliquots
1-  Put  the  end  of  the  DistriTip  into  a  suitable 

recipient vessel (according to the nature and 
volume of liquid to be dispensed).

Содержание DISTRIMAN

Страница 1: ...Other languages available on our website www gilson com...

Страница 2: ...eous volatile viscous or dense into many kinds of vial tubes microplates etc Distriman ensures reliable results in any domain and in all kindsoflaboratory forawidevarietyofapplications The positive di...

Страница 3: ...spare parts have been made available in case they are damaged as the result of extensive use 2 OPERATION Youshouldperformthefollowingsequenceofsteps 1 Select a DistriTip of suitable capacity 2 Fit the...

Страница 4: ...aliquots varies from 118 to 121 according to the type of DistriTip used Forexample theDistriTipMicroisidealfor llinga96 well microplate Thesyringecapacityis125 L sothe Distriman can dispense at least...

Страница 5: ...apacity of DistriTip tted The capacity of a DistriTip is marked on its barrel 2 Set the volume range by sliding the selector switch with your thumb to match the capacity of DistriTip tted The volume r...

Страница 6: ...82 mL 68 2 L 6 82 L Setting the volume 1 Pull up the black adjustment ring page 5 B 2 With your Distriman in one hand turn the adjustment ring with the other hand until the volumeter shows the require...

Страница 7: ...knobdownwards to its lowest possible position using your thumb 2 Immerse the end of the DistriTip into the liquid to be aspirated Whenaspirating ensurethatthe tip remains below the surface of the liqu...

Страница 8: ...on the dampened push button ensuring that each aliquot is com pletely dispensed and that the repeatabil ity is maximized see Recommendations Ejecting the Distritip You must change the syringe before a...

Страница 9: ...the side of the recipient vessel After changing the DistriTip or resetting the volume you may want to discard the rst aliquot Never try to force the volumeter beyond the recom mended volume ranges 4...

Страница 10: ...ply with the requirements of the ISO 8655 Standard bytypetesting Theadjustmentiscarried outunderstrictlyde nedandmonitoredconditions ISO 8655 6 Basis of adjustment Ex Reference temperature 20 C Relati...

Страница 11: ...y upwards 3 Make a note of the position and orientation of the spring K page 10 3 Pull on the retaining spring at an angle of 90 to the body of the Distriman This can be done by inserting a small scre...

Страница 12: ...ag You cannot remove the red ag E page 10 until you have separated the lling knob from the piston retaining assembly In fact the act of separating these two components will free the ag Toreplacethered...

Страница 13: ...2 Put all the parts into an ultrasonic bath for 15 minutes 3 Rinsethepartsthoroughlywithcleantapwater 4 Immerse all of the parts in a disinfectant bath for 15 minutes 5 Rinsethepartsthoroughlywithcle...

Страница 14: ...14 NOTES...

Страница 15: ...Devices GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France Tel 33 0 1 34 29 50 00 Fax 33 0 1 34 29 50 20 www gilson com Atika El Sayed Herv Ledorze General Manager Quality M...

Страница 16: ...3562 0027 USA Telephone 1 800 445 7661 or 1 608 836 1551 Fax 1 608 831 4451 Gilson SAS 19 avenue des Entrepreneurs B P 145 95400 Villiers le Bel France Telephone 33 1 34 29 50 00 Fax 33 1 34 29 50 20...

Отзывы: