6.
CONFIGURING WITH METSET
MaxiMet can be configured using Gill Instruments MetSet Software which is loaded on the CD supplied
with MaxiMet.
MetSet software can run on PCs running up to and including Windows 10 and can also be downloaded
from the Gill Website:
http://www.gillinstruments.com/main/software.html
For more details on all MaxiMet Parameters and their meaning refer to Appendices, Para 9.1.
6.1.
MaxiMet Default Configurations
GMX100 Factory Default Data String
NODE, TOTAL, PRECIPT, PRECIPI, TIME, VOLT, STATUS, CHECK.
┐
Q, 00000.200,000.200,2015-06-05T10:19:30.8,+05.1,0000,
└
36
Where
┐
STX
Q
Node Letter
00000.200
Precipitation Total
000.200
Precipitation Intensity
2015-06-05
Date
T10:19:30.8
Time
+05.1
Supply Voltage
0000
Status
└
ETX
36
Checksum
NOTES:
<STX> is the Start of String character (ASCII value 2).
<ETX> is the End of String character (ASCII value 3).
Checksum, the 2 digit Hex Checksum sum figure is calculated from the Exclusive OR of the bytes between
(and not including) the STX and ETX characters
The Status code will always read 0000 and has no functionality within a MaxiMet GMX100.
MetSet Default Settings
REPORT : NODE TOTAL PRECIPT PRECIPI TIME VOLT STATUS CHECK
PROTOCOL : GILL
PRECIPITATION SENSOR : ON
POWER : 0
COMMS : RS232
PRECIPITATION UNITS : MM
TIME : 2015-06-05T11:04:54
BAUD : 19200
NODE : Q
OUTFREQ : 1HZ
PUPMSG STATUS : ON
MSGMODE : CONT
PUPMSG REPORT : ON
ASCTERM : CRLF
PUPMSG UNITS : ON
ECHO : ON
MODBUS : RTU
MODADDR : 1
DATABITS : 8
STOPBITS : 1
PARITY : NONE
MODTERM : 10
MODICT : 1000