3 Performing measurements
3.3.3 Front panel buttons
P
-
9710 Switch On
Pressing the <on/light> key turns the P
-
9710 on.
P
-
9710 Switch Off
Pressing the <off> key turns off the P
-
9710.
Display Illumination On / Off
Pressing the <on/light> key (if the P
-
9710 is already turned on), switches the
display illumination on / off.
Measurement Result Display Update On / Off
Pressing the <run/stop> key stops updating the measurement display (the last
display will be frozen). The display‘s upper row will display "Stp" on the right
side. Pressing the <run/stop> again continues updating the display.
3.3.4
Menu description “Mode”
„CW“
displays the average measurement value for the pre
-
set integration time
(upper row). The lower row displays the selected wavelength or name of the
connected detector.
„CW Maximum“
displays the max. CW value input signal. Reset with <reset/
esc> key.
„CW Minimum“
displays the min. CW value input signal. Reset with <reset/
esc> key.
CW Maximum and CW Minimum should not be confused with Peak Values.
Peak Values will be calculated for each integration period (integration time) out
of the single ADC samples. CW Maximum and CW Minimum stores the largest
or smallest CW value since the last <reset> key press.
„Peak Maximum“
displays the max. ADC sample value of the integration peri-
od. It will be calculated new for each integration period (see menu "setup" /
"integration time").
„Peak Minimum“
displays the min. ADC sample value of the integration peri-
od. It will be calculated new for each integration period (see menu "setup" /
"integration time").
„Peak to Peak“
displays the difference of max. and the min. ADC sample val-
ue of the integration period. It will be calculated new for each integration period