3 Performing measurements
Use the <►> right and <◄> left arrow buttons to select the digit position
when changing numbers. The selected digit can be changed by using the
<▲> and <▼> arrow buttons. This also applies to the sign. Pressing the
<enter> button selects the adjusted number.
3.3.2 Menu Overview
The following table is a quick reference guide to the different menu items /
functions available in the P
-
9710.
menu item
submenu item
function
1. Mode
CW
displays the measurement respective of
any offset and calibration factors pro-
grammed
CW Maximum
displays the highest detected reading
CW Minimum
displays the lowest detected reading
Peak Maximum
shows the maximum detected peak level
Peak Minimum
displays the minimum detected peak level
Peak to Peak
displays the difference between the de-
tected peak maximum and peak minimum
level
I
displays the measurment result regarding
the distance between light source and
detector as luminious intensity / radiant
intensity
I
-
Effective (SC)
enables the measurement of pulsed light
signals with evaluation of the effective
intensity according to the form
-
factor
method defined by Schmidt
-
Clausen
I
-
Effective (BR)
enables the measurement of pulsed light
signals with evaluation of the effective
intensity according to the form
-
factor
method defined by Blondel
-
Rey