Manual 21568, Rev. F, March 2008
1
Introduction
1.1
Description
The 8035XA Series Peak Power Sensors perform true sample-based peak power
measurements on pulsed signals. The sensors are compatible with Giga-tronics Series 8540,
and the 8650 series Universal Power Meters and the Model 58542 VXIbus Universal Power
Meter (see Section 1.2.2). The sensors operate from 45 MHz to 18, 26.5, and 40 GHz. High
power versions of 5, 25 and 50 Watts, are available to 18 GHz. (see Table 1-1).
Peak Power sensors have three modes of operation: (1) CW, (2) Peak, internally triggered, and
(3) Peak, externally triggered. When operated in the peak modes, trigger-point to sample-point
delay (sample delay) is adjustable from -20 ns to 100 ms in 0.5 ns steps. Trigger levels are also
adjustable.
The Giga-tronics proprietary power sweep calibration system provides excellent linearity from
-20 dBm to +20 dBm in Peak modes, and from -30 dBm to +20 dBm in CW mode. Cal Factors
stored in EEPROMs in the power sensors automatically compensate for sensor frequency
response variations. This unique approach can be configured for automatic frequency
response correction. A detector output signal is provided for viewing the detected envelope of
the pulsed RF waveform on an oscilloscope. Use of a digital oscilloscope is recommended.
1.1.1
Accessories
Included:
3 each SMB (plug) to BNC (m) cables, 2 m (6 ft) long
3 each Cable Harness Wraps, 1.2 m (4 ft) long
Optional:
Option 02: 12 ft SMB (plug) to BNC cable
Option 03: SMB (plug) to SMA (jack) adapter
1.1.2
Product Returns
Should it be necessary to return the product to Giga-tronics, use the original shipping container.
If this is not possible, use a strong carton (350 lbs/in
2
bursting strength), or a wooden box.
Wrap the instrument in heavy paper or plastic before placing it in the shipping container.
Completely fill the areas on all sides of the instrument with packaging material, taking extra
precautions to protect the front and rear panels. Seal the package with strong tape or metal
bands. Mark the outside of the package “FRAGILE — DELICATE INSTRUMENT”.
If corresponding with the factory or the local Giga-tronics sales office regarding a product
return, please refer to the full model number and serial number. If the instrument is being
Содержание 80350A
Страница 4: ......
Страница 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Страница 13: ...Manual 21568 Rev F March 2008 ...
Страница 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Страница 17: ...Manual 21568 Rev F March 2008 ...
Страница 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Страница 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Страница 49: ...Figure 3 5 Digital Timing Diagram INT EXT Trig Mode Theory of Operation Manual 21568 Rev F March 2008 ...
Страница 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......