Series 8035XA Peak Power Sensors
4-6
Manual 21568, Rev. F, March 2008
0.1 nsec on the 8650A or 58542 power meters or by pressing the Enter button on the 8540C
power meter.
4.4.3
Delay Test
The Upper and Lower Limits in Table 4-8 are the minimum tolerances to test Delay
functionality.
NOTE:
The following test is suggested to verify sensor delay functionality. Sensor delay
accuracy is unspecified.
1.
Connect sensor to channel A of the power meter, and then calibrate the sensor. If an
80351A, 80352A or 80355A sensor is being tested, remove the attenuator.
2.
Select
PRESET
from the RECALL menu.
3.
Connect the sensor to a pulse generator with a 0 dBm, 50 MHz, 2 ms pulse.
4.
Use an oscilloscope or a frequency/width counter and measure the pulse width of the
SAMPLE DELAY output at the settings listed in Table 6. The sample delay measurements
in the chart are referenced at 1.7 V:
Table 6: Sample Delay Limits
Delay Setting
Lower Limit
Upper Limit
-20 ns
10 ns
30 ns
60 ns
40 ns
110 ns
1
µ
s
900 ns
1100 ns
10
µ
s
9
µ
s
11
µ
s
100
µ
s
90
µ
s
110
µ
s
1 ms
900
µ
s
1100
µ
s
100 ms
90 ms
110 ms
5.
Record the test results in Table 11 and circle the appropriate Pass/Fail test status.
This completes the Specification and Performance Verification Tests for the 8035XA Peak
Power Sensor. If the Sensor has performed as described in the preceding tests, it is functional
and correctly calibrated.
If the sensor fails to meet the criteria defined in these tests, refer to the Maintenance and
Troubleshooting chapters of this manual, or contact your local Giga-tronics Sales
Representative for assistance.
Содержание 80350A
Страница 4: ......
Страница 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Страница 13: ...Manual 21568 Rev F March 2008 ...
Страница 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Страница 17: ...Manual 21568 Rev F March 2008 ...
Страница 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Страница 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Страница 49: ...Figure 3 5 Digital Timing Diagram INT EXT Trig Mode Theory of Operation Manual 21568 Rev F March 2008 ...
Страница 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......